This is the second of two articles (Part I) which will provide suggestions on how to improve
the HPLC (UHPLC) method development process. - PART II
INITIAL
METHOD DEVELOPMENT:
- Before
you start, learn what you can about your sample, its hazards, solubility
and properties. Conduct a quick literature and/or keyword search on the
web using a popular search engine (e.g. GOOGLE). You can often find many
journal articles, white papers, application notes and chemical data on the
web in just a few minutes.
- Determine
which liquids your compound(s) are soluble in. Use a pipette and several glass vials
or tubes with different solutions (pH is important too). This will narrow
down the types of mobile phase and chromatography modes that you can use.
- Column choice is the most important part of method development. The stationary phase that you choose has the single greatest effect on selectivity! Select
the right column and mode of chromatography. Most RP methods should start with at least a modern ultra high
purity, metal free type B silica column with a C8 or C18 support (*But you must select the column type that best suits your application). For
small molecules (<1,000 Daltons) select a standard sized analytical
column with a 2.5 to 5.0 micron particle size and pore
size between 60 and 120 Ã… (e.g. 4.6 x 150 mm; 3.0 x 100 mm). *For larger peptide or protein molecules you will need a particle
with a large pore
size (~ 300 Ã…). For optimal results, columns with very small internal volumes should be paired with HPLC systems with similar ultra-low internal delay
volumes. Your HPLC system should be optimized to balance the needs of good
mixing,
low delay volume and proper sampling
rate for your method. If the method is likely to be transferred to different
types of HPLC systems, then you may want to initially stay away from the
< 2.1 u particle supports for your method development. At this time, these tiny
particles can not be reliably packed into columns as well as the larger sized
particles and generally have much poorer %RSD values than larger particles
(i.e. 2.5 to 5 u). Method development is initially easier & generally more rugged
using conventional particle sizes [Keep things simple when starting out. You can always change later on]. Once you have selected a column and
decided on a flow rate,
make sure you calculate
and measure the actual column void volume to find the 'Tzero value' (column dwell volume). You will
need this value to find out if your compound has been retained
on the column (K
prime) and to also determine most of the needed performance
calculations and parameters.
- Once you
have selected an HPLC column (if possible, please start with a brand new column) you
will need to test it and establish a baseline to show that it meets both
the manufacturer’s specifications and, far more importantly, that it meets
your method’s requirements. In most cases, do not use the manufacturer’s
QC test solution for this. Those stds are designed to allow the column to easily
pass manufacturing QC, not the more critical requirements that you may
need to prove. We prefer to use a real sample mixture (~2 compounds) that
are similar to the type proposed for the method. They must be well
retained on the column (K
prime of > 2), easy to prepare, stable and reliable. This std and
the specific method you develop for it, will be used to prove the column’s
performance (i.e. R,
S, K prime, Tzero,
Plates).
It will be used again, when the column’s performance is called into
question.
- Notes on Mobile phase and additives: Keep it simple. Avoid the use of any additives such as ion-pairing reagents when first starting out. They are overused in general and can cause problems later on. If required, they can always be added later on. Use only HPLC grade solvents, fresh RO HPLC grade Water and the highest purity acids, bases and/or additives, if required. Use only filtered (0.22u) products. Always dissolve samples in the mobile phase (or a weaker solution) for injection. For many RP methods a low pH mobile phase (~ pH 2.5) provides a good starting place. Samples with a pKa far enough away from this value are likely to be retained, stable and unaffected by small changes in pH. *pH is usually one of the last parameters which is optimized.
- Use a
Gradient Method to find the approximate elution conditions of the mixture AND
make sure you are detecting all of the compounds injected onto the column
[For dedicated RP isocratic methods, start with a high organic mobile
phase % (i.e. 95%) and record the results. Continue to reduce the organic
content in steps of 10% and observe where the best compromise exists
between retention and elution]. For RP gradient methods, start with a very
high aqueous % (e.g. 98%) and run your gradient, slowly, to a very high
organic % (e.g. 95% or 98%, not 100%!) to make sure you retain, hold,
and then elute everything. Your mobile phase conditions must be
strong enough to elute everything off the column during the gradient
portion of the run (long enough hold time). Please Do NOT include the
gradient reversal portion back to initial conditions (wash and
re-equilibration) of the gradient as part of the same analysis method. The
method should end after the gradient has reached and been held at its
maximum level for a period of time (we refer to this as the "hold
time"). It should NOT switch back to the initial conditions to
re-equilibrate the column. Don't make this novice mistake. Column flushing
and re-equilibration should be a separate method and/or step, separate
from your analysis method. If you include your column flushing
(washing) and re-equilibration steps as part of your analysis method,
you are also forcing the baseline's slope to change radically. This may
interfere with integration plus include extra peaks and baseline changes
that you are going to have to integrate, identify and explain to others
(e.g. auditors). Additionally, including these wash steps as part of the
same method wastes time as you must wait for these steps to complete
before you can start to process the data obtained in the analysis method. Summary:
Develop methods like the professionals do and create separate
Analysis and Flush/Re-Equilibration Methods (or steps).
- If your
compounds can be seen with a UV/VIS detector, then make sure you are using
a Diode Array Detector (aka, DAD or PDA) set to scan a wide range of
wavelengths (e.g. 210 - 410 nm). Method development should not be
performed using a single or multi-wavelength detector. This invites
errors, limits the utility of the method and does not result in any cost
savings. You must have a full scanning detector so you can detect all
possible peaks at the same time. If you use a single or dual wavelength detector
you may not know if an impurity has been introduced or if you have a
co-eluter (because you will have no way to detect it). Generic starting
settings for Wavelength
Bandwidth should be ~ 8 nm, software based Reference
Wavelength OFF and the sampling
rate must be set to collect at least 20+ peaks/sample peak of the
narrowest sample peak observed and integrated (at ½ height). Run with
scanning turned 'on' for all analysis methods and review the spectral data
for each run. This provides important qualitative data about the compounds
which may be used for purity determination and also to demonstrate how the
method is selective for the compound(s) of interest. *If your compounds do
not absorb well (weak chromophores), then you may still want to have a
UV/VIS detector inline (first) with a secondary detector second. The
UV/VIS detector will still be very useful for troubleshooting and
detecting other compounds. Select an appropriate type of detector and
compatible mobile phase as required for the secondary detector (e.g. RID,
EC, FLD, ELSD, CAD, MS...). Note: If available, LC-DAD-MS is one of the
most useful instrument setups for LC method development.
- This needs to be repeated (from Part I).... Before starting ANY HPLC analysis, the HPLC pump must be running and operating with no problems, achieving a stable baseline, steady flow rate with as little pulsation as possible (~ 1% ripple or less). Accuracy depends on this. Do not begin any HPLC analysis unless the HPLC pumping system is working perfectly.