Translator for HPLC HINTS and TIPS for Chromatographers

Showing posts with label Void. Show all posts
Showing posts with label Void. Show all posts

Saturday, February 4, 2017

Determine the HPLC System Dwell Volume (Gradient Delay Volume)



Note: The total HPLC gradient system dwell volume is different than the HPLC column’s void volume. Two different terms for two very different measurements.

When we perform gradient HPLC analysis, the mobile phase composition is changed over a period of time. The mobile phase is mixed in real time by the pump(s), mixer and/or valves, then transported to the injector and finally, on to the head of the HPLC column. The total volume of liquid contained between where the mobile phase is mixed and the head of the column helps us determine when the newly mixed solution arrives at the column head (it is not instantaneous). This delay is often referred to as the gradient delay time (or delay volume) and its value will vary for different HPLC systems due mainly to differences in tubing dimensions used, pumping system type and the design of the flow path. 

For example: If the system dwell volume is found to be 1 ml and the flow rate used is 1.000 ml/min, then the gradient delay time is one minute. 

So how do we know what the system dwell volume or gradient delay volume is? Well, we measure it of course!

Measure the ‘System Dwell Volume’ (aka: Gradient Delay Volume)*:
(1) REMOVE any HPLC column(s) and install a Zero Dead Volume Union (*ZDV) or a restriction capillary of know volume in its place.
(2) Prepare Two Different Mobile phase solutions:
Bottle ‘A’: HPLC grade Methanol (MeOH).
Bottle ‘B’: HPLC grade Methanol with 0.1% acetone added (v/v).
(3) Set your UV/VIS detector to 265 nm (8 nm Bandwidth, Reference OFF).
(4) Program a suitable system flow rate and create a simple Gradient Method (linear change) which starts at 0.0 minutes with 100% ‘A’ (HPLC grade Methanol) and 0% B (HPLC grade Methanol with 0.1% acetone added) and runs to 0% ‘A’ and 100% ‘B’ for about 10.0 minutes (actual times used will depend on your selected flow rate).
(5) Flush and degas both solutions, ‘B’ first, then ‘A’ through the system until you get a nice clean, flat baseline. Make sure their is enough backpressure on the pump (>40 bars) to obtain a stable signal (use a restrictor or back-pressure regulator if needed).
(6) No injection should occur during this method.
(7) Start the method (RUN) and observe the 265 nm signal over time. At some point you should observe the signal begin to rise. When you see this signal change occur, the acetone has finally made it from the pump head to the detector’s flow cell. Make note of the time this occurs. 

Using the known flow rate and observed signal change time, you can now estimate the total system dwell volume. 

Example: If you observe the signal start to rise steeply at 2.00 minutes and your flow rate was 1.000 ml/min. Your system dwell volume would be 2.000 mls. 

A more accurate system dwell volume value can be obtained by next running the same method with an injection of acetone (e.g. 1 ul) and noting the time at which the injection peak is first seen. That will give you the time it takes the sample (and therefore the volume needed) to go from the injector to the flow cell. If you subtract this time off the system dwell time you recorded in the last test, you will have the actual measured time from the pump head (or proportioning valve) to the head of the column (vs the flow cell). Normally the volume contained in this tubing and flow cell are very small relative to the volume in the rest of the system, so we can ignore them. However, when using some of the very low volume columns (e.g. 2.1 x 50 mm), the volume contained in these areas can become significant so when appropriate, we need to be aware of them.

Failure to take into account changes in HPLC system dwell volumes can result in methods which no longer work or provide different results. This is because the gradient rate change you program in your method may not allow enough time for the new mobile phase composition to reach and flow all the way through the column in the time that you have programmed. A common mistake we see is when users forget to adjust the gradient profile when changing column dimensions or program changes using too fast a time.

BTW: One common trick we use to improve compatibility between systems which have different dwell volumes is to include an initial (time 0.0)  isocratic hold-time into the start of each method. If all systems used have system delay volumes under 3 mls, then add a 3 minute isocratic hold time at the start of each method (if 1.000 ml/min flow rates are used), before any gradient starts. While not the best way to deal with the issue, this type of “cheat” can make it possible to quickly adapt a method for use on several different system types.

*Note: This is a generic method to determine the system dwell volume or gradient delay volume. Detector signal buffering and flow cell volume also adds to the delay and in some cases, must also be accounted for too. There are many other methods which can be used for this determination as well. This proposed example serves to illustrate the concept only.

Saturday, September 24, 2016

HPLC Peak Splitting. Common Reasons For It



True "Split" HPLC peaks, not resulting from co-elution of another peak, can be caused by a number of chromatography problems. Here are a few examples and their solutions:

  1. Sample overload. Sample overloading is one of the most common reasons for observing peak "splitting". Reduce the sample concentration by factors of ten to see if the peak shape improves. 
  2.  A poor quality HPLC method. Poor quality methods which do not use mobile phase solutions which are at an appropriate pH (*If the pH of the mobile phase is close to the pKa of the sample, then split peaks may result); which does not dissolve the sample in (should be fully soluble) or are unstable, show sample or mobile phase precipitation can cause this effect. Always check solubility before starting.
  3. A partially plugged or fouled column. A dirty or fouled column (from not washing down properly with a solution which is STRONGER than the mobile phase). Analysis methods should be followed by separate wash methods to remove all bound material and any late eluters,
  4. Wrong injection solution. Peak splitting may be the result of dissolving and injecting your sample in a solution that is stronger than your mobile phase. Dissolve and inject samples in the mobile phase or in a solution which is a slightly weaker solution (not stronger).
  5. A poorly packed column, void at column inlet, a dirty frit or poor mechanical connection (i.e. improperly swaged fitting). These types of structural or mechanical defects can each result in peak "splitting" (all of these are less common today than in the past using modern HPLC columns). When present, a dirty inlet frit can be replaced with a new one, or the column can sometimes be backflushed to remove any accumulated material. Connections should always be double checked.
  6. Detector data rate set too low. Too few peaks collected over time may result in integration errors and inaccurate peak symmetry problems. Read more about how to determine the best data collection rate at this link.

Saturday, January 30, 2016

HPLC Column Support Pore VOLUME

If an HPLC column had no packing material inside it, then the volume of liquid contained in the cylinder could be calculated using the formula for the volume of a cylinder as follows: 



      Volume of Cylinder = Pi * r2 * L;     
          [where Volume is in ul; Pi = 3.14; r = column radius (mm) and L= column length (mm)]
  Example: Using the above formula, a 4.6 mm x 250 mm column would have an empty volume of 4,155 ul (~ 4.16 mls).

For most chromatography applications we pack the column with a high surface area porous media. Often this is a silica based support. This support media fills the empty space inside the column reducing the total volume accessible by a liquid (or to the samples). If the media used was not porous, it would fill most of the space (depends on size and shape of media). Most commonly used chromatography supports are porous and leave about 70% (0.7) of the original volume available to the mobile phase and sample [Pore Volume = Surface Area (m²/g) x Pore Diameter  (Å) / 40,000]. Based on this information, we use a value of 0.7 as the average pore volume for a packed chromatography column (some supports will have pore volumes which are larger or smaller than this value. The manufacturer will often measure it and provide the value on their published specification sheet).


Using a typical 4.6mm x 250mm column we found the total volume to be 4,155 ul (4.16 mLs). If we now multiply this empty column volume by 0.7 (note: use 0.7 or 70% for columns with fully porous particles and 0.55 or 55% for superficially porous particles) we obtain 2,908ul total volume (2.9 mLs). This is the estimated volume of the fully packed column. This value is very important as it provides an estimate of what the column dead volume will be so we can calculate the 'T' zero time of an unretained analyte. This estimate will depend on the column dimensions, using our HPLC method (be sure and take into account the measured flow rate to determine the column "dead time"). This is one of the very first calculations you make when starting or modifying an HPLC method and is critical information to know at all stages of method development. All chromatographers should know how to estimate this value before using an HPLC system. *You should confirm this estimate by injecting an unretained sample onto the column and measure the retention volume, then compare the two values. The measured value is the most important number (the one we use for calculations), but the estimate should be close (+/- 15%). The estimate is still useful for troubelshooting and method development as when combined with K prime, it provides a quick measure if chromatography has occurred (retention).

For more information on the importance of knowing the HPLC Column Dead Time, please refer to this article link

Notes: The measured support pore Diameter (SIZE) is important for determining if the sample will have access to the inside of the support (e.g. A support with a pore size of 80Å will be too small for most large peptides or proteins, but a support that is 300Å will allow access to many, not all, larger molecules). A support with too small a pore diameter will not allow the sample to access the high surface area inside the support. Instead, the sample will be unretained and pass by it eluting at the column's void volume. This is the basis of SEC or GPC analysis where we use columns with different pore sizes to "filter" samples based on size. Large pores for large Mw samples and small pores for low Mw samples. A general rule is use 300Å or larger pores for samples with Mw > 10,000 and 80Å to 150Å for smaller samples.

More info on pore volume can be found at this article link: https://hplctips.blogspot.com/2014/12/hplc-column-pore-volume-or-pore.html