HPLC column temperature is a critical variable that we adjust and optimize during method development. We use it as a variable
during the method development process to improve solubility, optimize peak shape and increase
resolution. Once established, it must be carefully controlled during the
method analysis to provide reliable and reproducible analysis results. Change
the column temperature and you may also change the results obtained. This is a fundamental method development tool and must not be forgotten.
If you are developing a new UHPLC method OR perhaps scaling an HPLC method to utilize 2.5 micron or smaller support particles, then you may observe a loss of resolution or poor peak shape in the new method. There are many reasons why this may occur, and the most common ones relate to not optimizing all of the method parameters correctly when scaling the method (e.g. dwell volume too large, flow cell volume too large, injection volume too large, sample rate too slow, flow rate not optimized, mobile phase composition changes not in scale with the gradient...). But there is another reason...
Resolution may be reduced or lost when all of the initial scaling and instrument set-up parameters are optimized. What is the most likely reason for this? In many cases the use of substantially higher flow rates (relative to linear flow rates) and the use of smaller diameter particles results in much higher backpressures (you may recall that if you halve the particle size, the backpressure increases 4x). The resulting backpressure might be 2, 3 or even 4 times higher than observed in the original method. While these higher backpressures were well within the operating parameters of the HPLC system used, the results obtained were poor. The possible cause? The much higher backpressure increased the amount of frictional heating inside the column, raising the actual analysis method temperature and changing the separation conditions.
Pushing mobile phase (liquid) through a chromatography column generates heat and pressure. The heat generated increases the actual temperature of the column and reduces the viscosity of the fluid. In conventional columns (i.e. 4.6 x 150 mm, 5u) at 1.00 ml/min, this heating effect is minimal, but at much greater column pressures, > 400 bars, the frictional effects may be substantial. These types of very high pressures may be seen with methods which utilize columns containing the smallest particles (1.9 to 2.5 micron). Enough to change the temperature in the column by several degrees (e.g. >5 degrees C) and result in different method conditions. So, what can you do about this? The most direct way to address the problem is to run the same method at a lower temperature (perhaps decrease by 5 C to start with). This will slightly raise the backpressure (lower temperature equals higher viscosity), but it should cool the column and restore the original temperature conditions used. Additionally, we suggest that you always start column equilibration using a flow ramp to gradually increase the flow over time and reduce the overall heating effect and resulting "shock" placed on the column. An initial delay at equilibration may help reduce these effects (gradually ramp up to the regular flow rate and hold). You may need to try several temperatures and this may be easiest to do if your HPLC has a column compartment with heating and COOLING capabilities. Optimizing the temperature and internal pressures may increase the column lifetime and result in better overall data reproducibility.
The HPLC "Hints and Tips" found here will be of interest to anyone performing HPLC, UHPLC (UPLC), LC/MS, SFC, CE or GC analysis. Many of these free tips, taken from professional classes that I teach, discuss fundamental to advanced level chromatography concepts in a practical way. Occasionally, I will also include laboratory computer related tips here too. Please "follow" this blog to receive updates of new posts. Hint: Type keywords into the SEARCH bar to quickly find other articles.
Translator for HPLC HINTS and TIPS for Chromatographers
Saturday, January 6, 2018
UHPLC TIP: Reducing the Column Temperature to Offset Frictional Heating Effects (Causing Poor Resolution)
Labels:
Back Pressure,
Backpressure,
Column,
Column Oven,
Column Temperature,
Frictional heating,
HPLC,
HPLC Tips,
Method Development,
Particle,
Resolution,
Scaling,
Shape,
Temperature,
UHPLC
Saturday, December 9, 2017
Evaporative HPLC Detectors; CAD (Charged Aerosol Detector) and ELSD (Evaporative Light-Scattering Detector)
- If you wish to read about their development and/or operating principles, then please review the early published patents and many articles available through the web. Be cautious when reviewing any "sales" brochures or articles on these detectors as a great deal of misinformation may be found.
Applications: CAD
and ELSD are both used with a wide range of non-volatile
sample types. Targeted at compounds which have weak or no UV chromophore (e.g. Carbohydrates, fats, lipids, triglycerides,
polymers, surfactants, oils).
Thousands of application notes and journal articles are
available for both types of detectors (esp. for ELSD with almost 40 years of
use) and a keyword search on the web is the best way to find them. As someone who was
involved in the early development and design of these detectors, I have used them successfully to develop several hundred different types of methods. They have proven to be useful for a number of difficult applications, but their higher cost and even higher training and skill requirements still place them outside of most users labs. As with LC/MS
detectors, CAD/ELSD modules may require far more maintenance and advanced training to use
than most chromatographers have received. As such, it is my opinion that you
consider their potential use in your projects only after other
more conventional methods have failed to provide results. Due to the high level of training needed, difficulty to operate and maintain, high cost of operation and poor reproducibility, IMHO they should be a "last choice".
Detection: NOT “Universal” detectors (sourced to
marketing misinformation from vendors and early academic reviews which over simplified their 'operation', not of the actual commercial instruments). While detection is partially based on the analyte’s chemical or
physical properties, the actual output observed is in fact also based on the properties of the mobile phase (volatility and purity), sample volatility/stability and
especially the many different custom detection settings chosen by the user (gas
flow, heater temperatures, flow rate, specific detector used, level of contamination
inside the detector). As such, their output is very subjective since it is
based on both the specific chromatography method selected, the condition of the
detector, the lab environment used-in, and the detailed operational settings chosen by the operator. They can detect
everything from dirt, buffers, undissolved chemicals or particulate matter in your
mobile phase. Even pressure changes on the detector's exhaust line can effect the output.
“Destructive” Detectors:
As with an LC/MS system, the mobile phase is evaporated away from the
sample and sample collection is not possible at the exhaust. They are best used as a secondary detector, with a primary detector sch as a UV/VIS module placed
in front of the CAD/ELSD (to increase your chances of
detecting something that the CAD or ELSD may miss). ELSD and CAD units will NOT detect all samples. If sample collection is required, a low volume, micrometer valve flow-splitter can be fitted to the evaporative detector’s
inlet port. Note: Depending on the flow-splitter's split ratio, the detector’s signal output may be
reduced.
Mobile Phase
Requirements: Evaporative detectors require a fully volatile mobile phase (similar to LC/MS
requirements). The use of non-volatile additives can contaminate or damage them
(no phosphate buffers!). Use of non-volatile buffers or additives, low purity materials, contamination of the gas, mobile phase or by samples may result in excessive noise levels limiting detection. Use high-purity grades of mobile phase and additives. Examples of Mobile phases used: "Popular LC/MS and HPLC Volatile Mobile Phase Buffers"
Isocratic and
Gradient Capable: Unlike RID or EC, CAD/ELSD allows the use of gradients and the use of UV obscuring solvents.
Because the mobile phase is evaporated away, little to no
baseline drift occurs during gradient analysis (often improving integration results). Sample types which dissolve best in solvents such as
methylene chloride, acetone, chloroform or other strong UV absorbing solvents may find that
these detectors assist in developing better quality methods. Reduced gradient baseline drift plus the option of using UV absorbing solvents are two characteristics which make them well suited to application areas such as lipids, polymers and oils. Flatter baselines allows for better quality peak integration.
Gas Requirements:
Similar to the requirements of an electrospray LC/MS system, both CAD and ELSD
modules use very large volumes of high-purity gas (i.e. Nitrogen) to safely evaporate
the mobile phase away. Exhausting these large volumes of solvent vapor and gas into a fume hood is just as important during site prep. Be sure and factor these costs and the required space into any site-prep plan.
Operational Reproducibility
and Method Transfer: Recording
the exact detector settings used in the method may not provide any guarantee of
being able to duplicate the results obtained. No two instrument models are the same
so results may vary (similar to LC/MS). Results obtained for each sample are relative to the specific instrument, the chosen settings & method used (again, much like LC/MS) and the internal condition of the detector used. Compare the many critical heat, gas flow and atomization
related CAD/ELSD settings to the more common UV/VIS detector where only the
wavelength, bandwidth and flow cell dimensions need to be specified to easily duplicate
the detector setup. CAD/ELSD internal contamination levels, nebulizer spray patterns, gas
flow rates, quality of the mobile phase and operator training may all contribute
to variations. *As with all methods and detection systems, proper training and good method design will insure success.
Quantitation: Can
be used for quantitative analysis across a wide dynamic range spanning multiple
orders of magnitude with some success. High quality reproducible methods are achievable with both types of detectors, but will require calibration tables with many additional standards per order of magnitude.
Linearity and Output Characteristics:
Except in the most narrow concentration ranges, neither detector is likely to provide
a linear response. Different samples will need their own full calibration table and curve fit, per method. Quantitation can be improved through the use of larger numbers of calibration
levels (more than normal) plus a high quality chromatography data analysis software package which includes
many available non-linear curve fit options (polynomial, quadratic, sigmoidal, exponential,
log…etc). Output often changes across orders of magnitude so be sure to optimize the curve fit for each sample type. Different sample types will often have different response outputs at different retention times. This is most easily observed during a gradient analysis. As the mobile phase composition changes, so does the response for EACH sample (this is NOT a UV/VIS detector).
Optimization Process:Unlike a UV/VIS or RID system which simply needs to warm up and stabilize, CAD
and ELSD systems may require a methodical optimization process of adjusting the flow rate, gas
flow and heating temperatures to optimize the measured S/N peak ratios for each sample and each method used (yes, ever one of them). Optimization of
detection conditions usually involves making multiple measurements (Peak and Baseline S/N
ratios) to find the best settings to use with each sample type and method. This optimization process is time consuming and changes may need to be made to the method over time as the detector fills up with baked-on sample material (changing the spray pattern via nebulization changes).
Operational Complexity:
Methods which utilize CAD/ELSD systems may be more complicated and time
consuming to learn, use and validate then conventional detectors. Specialized detector
cleaning procedures are often needed. The detectors may become internally contaminated during use (sample builds up inside the unit). Failure to clean and maintain them may lead to high noise levels and/or inaccurate results. Due to the additional maintenance needs, lack of traditional linearity, and overall complexity, we recommend their use only when: (1) Conventional
detectors or methods of analysis are not possible or unsatisfactory and (2) where the operator has demonstrated a high level of practical hands-on training through use of the detector and/or has sufficient experience
(advanced level) in chromatography.
For more information:
- “Advanced Evaporative Light-Scattering Detector (ELSD) For A Variety of HPLC Applications”;
- “Evaporative Light-Scattering Detector (ELSD) Optimization Procedure. Varex ELSD IIA, Technical Note # 0024”;
Saturday, November 4, 2017
Repair Missing Or Corrupted Windows System Files Using the System File Checker Tool
It has been awhile since I posted any Microsoft Windows software tools or tips so here is a MS utility program. Most laboratory instruments operate under MS Windows software control and you will be able to keep things running smoothly if you take the time to learn and use many of the available utility programs.
In addition to the very useful "Restore Point" feature found in Windows (discussed in an earlier post, which is great for solving a number of issues), Microsoft has another time-saving utility tool built into most versions of Windows (i.e. XP, Vista, 7, 8 and 10). The Windows System File Checker (simply known as, SFC) offers users the option to scan their Windows operating system files for corruptions AND restore any found corrupted files, all automatically! Running the utility is very easy. First, make sure your account has Administrator privileges, then use the Run Command prompt to run the program SFC 'As Administrator' *Type: sfc /scannow
More information about this useful utility can be found on Microsoft's website support page.
- Note: Before using any software utility program, make sure you first have permission and authorization to do so. Back up your system programs and any data files. Create a Restore Point to protect the basic settings too. Take precautions before using any utility programs and do so at your own risk.
In addition to the very useful "Restore Point" feature found in Windows (discussed in an earlier post, which is great for solving a number of issues), Microsoft has another time-saving utility tool built into most versions of Windows (i.e. XP, Vista, 7, 8 and 10). The Windows System File Checker (simply known as, SFC) offers users the option to scan their Windows operating system files for corruptions AND restore any found corrupted files, all automatically! Running the utility is very easy. First, make sure your account has Administrator privileges, then use the Run Command prompt to run the program SFC 'As Administrator' *Type: sfc /scannow
More information about this useful utility can be found on Microsoft's website support page.
Saturday, October 7, 2017
Preparation of Phosphate Buffered Saline (PBS)
PBS
While not commonly used in liquid chromatography, PBS
solution is commonly used in preparing samples. Because chemical forms and grades vary, there are several recipes on the web. By popular request, I am
provided a common laboratory recipe for a 1X PBS solution here.
To make Phosphate Buffered
Saline (PBS), 1x pH 7.4 solution:
Method:
1. To a 1-liter Beaker, add the
following four anhydrous salts:
a. 200 mg KCI
b. 8,000 mg NaCI
c. 240 mg KH2PO4
d. 1,400 mg Na2HPO4
2. Add about 850 ml of deionized
water and stir to dissolve the salts. Adjust the pH to 7.4 with 10% phosphoric acid (phosphate
solutions should be adjusted with phosphoric acid only). Transfer to a 1 L volumetric flask (class-A). Fill to the “line”
with more deionized water. Cap and mix a final time to insure a uniformly mixed solution.
3. Filter
the final solution through a suitable 0.22 micron filter before use.
Note: pH increases with decreasing temperature.
Subscribe to:
Comments (Atom)