Translator for HPLC HINTS and TIPS for Chromatographers

Saturday, January 25, 2014

Diagnosing & Troubleshooting HPLC Pressure Fluctuation Problems (Unstable Baseline)

Few things in chromatography are more frustrating than dealing with large pressure fluctuations (>1% ripple). If the pump pressure is unstable, and fluctuating up and down, then it will negatively impact your ability to analyze, measure and integrate sample peaks in a reliable manner. A smooth, flat baseline is needed to run and develop methods, collect the data (peaks), integrate and report the results which are reproducible. Baseline instability during an analysis may lead to the entire analysis being declared invalid.

So what causes the HPLC pressure to sometimes fluctuate in a wild manner up and down on your HPLC system? Unfortunately, many things... Most result from poor training, incorrect operation techniques, but some are maintenance related so be sure and keep your chromatograph in excellent condition. Maintain a logbook for each instrument and record what types of maintenance and service have been performed over-time, with the date and list of parts used/replaced. Additionally, maintain a preventative maintenance schedule (e.g. every six months) to inspect and clean the entire HPLC system to check condition, verify operation and minimize unproductive down time. 

HPLC Pump or System Pressure Fluctuation Causes and Solutions:

  • Air / gas In the Liquid or Mobile phase (Failure to Degas Mobile phase OR loose fittings) --- Air gets into the system due to a leak or from gas trapped in the mobile phase. Find and correct the cause of the leak and/or degas the mobile phase (use continuous Vacuum degassing or a Helium sparging system only). Leaks are the most common cause of instability, but insufficiently degassed solution is a close second. Make sure your degasser is working 100% correctly (they require professional servicing every 5 years). HPLC pumps require degassed mobile phase for reliable operation.
  • Loss of Prime. Improper Priming of the System --- Failure to flush ALL of the lines with freshly degassed mobile phase, before use (every day), will often result in all kinds of instability problems until all of the old gas-filled mobile phase has bee purged from the system. *This could take many column volumes of liquid. Make sure you account for any vacuum chamber volume too. Properly prime the pump heads before use.
  • Sticking Check Valve(s) --- If air is exiting the pump outlet, the pump will not function properly. Both Inlet and Outlet valves should be inspected. Remove and clean the check valve(s). Be sure the pump is fully primed with liquid as the check valve might just have an air bubble in it (common on Waters, Thermo and Shimadzu systems). Sometimes sonication of the valve for ten minutes in a beaker containing warm solvent does the trick (e.g. MeOH or IPA/Water). Though very rare, ACN has a bad reputation for polymerizing in solution. If the system has sat unused for a long time OR was not properly flushed out when last used, it is possible that particulate matter may clog the flow path. Small sticky particles may form (ACN polymerization) and cause the check valve to stick inside the housing (use fresh, filtered solvents only to prevent these problems). Clean and inspect any suspect valve first. Replacement of the check valve may be needed in some cases to restore operation. Note, this problem of "sticking" check valves is most likely to be an issue in HPLC pumps with mechanical (gravity or spring) check valves with ruby balls, not modern style active inlet check valves ("AIV") which are electromechanical (solenoid valves) and are very reliable, much less susceptible to these problems. In any case, verify operation of all valves while under pressure (backpressure is needed for them to function correctly).
  • Worn Pump Piston Seals --- Commonly observed as rapid up/down spiking on all channels and an inability to maintain or produce backpressure (the pump will often prime with no problem, as this is done at low-pressure). Run a formal pump high-pressure leak test at max pressure to confirm (remove the column and replace with a calibrated backpressure restriction line for all testing). Clean pistons and replace piston seals to repair (you should have spare pistons and seals on hand). *Seals are a maintenance item so expect them to wear out and need regular replacement.
  • Flow rate too low (may be inappropriate for system). Running at a flow rate that is below the optimum range of the specific instrument (i.e. System rated for 200 to 2,000 uL/min, but run at 100 to 200 uL/min or at the limit of the range) may result in an unstable baseline. The cause may be due to pump cavitation, loss of prime, non-optimized piston stroke volume.
  • HPLC System Back-pressure too low to maintain prime in system. Most types of analytical HPLC systems require a minimum system back-pressure of 40 or more bars to maintain enough pressure (mechanical compression) on the component parts to run in a reliable fashion (*Water's Article number: 32564 states the back-pressure must be at least 1000 psi for their Alliance systems). Too low a pressure often results in a loss of prime, cavitation, mixing problems, turbulence and poor reproducibility. Correct sizing of column, particle size, flow rate and mobile phase composition should all take into account achieving enough back-pressure on the system to maintain a stable baseline throughout the entire analysis. Monitor the system back-pressure at all times for stability. High quality research grade HPLC systems are often capable of maintaining stable isocratic flows with less than 1% ripple and 0.2% ripple common ("ripple" is a term we often use to describe the pump's pressure output over time relative to the baseline (S/N)).
  • Mixing Problem (gradient or isocratic online mixing) --- If your active mixer or proportioning valve (AKA: Gradient valve) is defective or dirty, then one or more of your mobile phase channels may not be getting to the pump. Air would most likely be mixing with the mobile phase causing the unstable flow. Clean or replace the valve. Note: Always try flushing the gradient valve with pure IPA, then DH20 for about twenty minutes. This sometimes restores operation by wetting and flushing the internal seals (which may dry out).
  • Wrong Pump Solvent Compressibility Settings --- In HPLC we routinely subject different liquids to very high pressures which can result in measurable liquid compression. The degree of actual compression for each liquid varies, but the modern HPLC pump can compensate for this to improve the accuracy of the mixing and flow delivery.  Most pumps provide for user adjustable solvent compressibility values. If the value input varies a great deal from the actual liquid in the system, then it can result in pressure fluctuations. Example: Water has a value of 46, but Methanol 120. Using the wrong value can cause instability.  
  • Poor Solubility, Mobile Phase  --- Sometimes the mobile phase which has been prepared (or mixed online) is not 100% soluble. This could be due to an inorganic salt additive which has not gone into solution or failure to fully mix and filter the mobile phase before use. Ultrasonication, a bit of heat and stirring for 20 minutes can help to get everything dissolved. 
  • Dirty inline filter --- A fouled or partially plugged filter can disrupt the normally smooth flow into a turbulent one. Some are installed as part of the pump (i.e. HP/Agilent brand pumps) and should be changed out every month (Yes, for the PTFE frit, replace it once a month with a new one). Other systems use these pre-filters downstream of the pump before the injector. Clean or replace all filters frequently. If used in your system, these are regular maintenance items and should be part of a general 'PM' program.
  • Dirty Solvent Pickup Inlet Filters: These can become obstructed or fouled over time (esp. if used with aqueous solutions!). Just as with any built-in filter, the multiple solvent inlet pickup filters should be cleaned or replaced on a regular basis to prevent particulate or any material which may contaminate or restrict the flow path from entering the system. Mobile phase pickup filters are often 10 to 20 um and connect to the bottom of the low pressure (e.g. Teflon) solvent lines in each bottle. If you use 316 Stainless steel filter (recommended for organic solvents), they should be removed, cleaned in an ultrasonic bath, rinsed and replaced monthly. If you use sintered glass or other disposable type filters (often used with aqueous solutions), they should be disposed of on a regular basis and replaced with new ones (replacement, not cleaning is recommended because sintered glass can not be sonicated and should be disposed of to prevent bacterial, mold or fungal contamination). A quick way to check if one filter is causing the pressure to fluctuate is to remove the filter from the one line, then re-test the system. If the problem goes away, then returns when you re-install the filter back on the line, the filter may be obstructed (replace it),
The above list includes some of the most common reasons for unstable baselines. Other non-pump related causes would include a bad / old detector lamp(s) or contaminated mobile phase. To find the cause, test and verify the operation of each component part of the HPLC. Troubleshooting Advice: Test one part at-a-time, before moving to the next part. Never assume anything, test, re-test and verify or prove at each step.

Saturday, December 21, 2013

Two Common HPLC Problems and their Causes (Sudden changes to either the HPLC Backpressure or Peak Shape)

   Let's take a quick look at two different problems which you may encounter when operating an HPLC system. We start with the basic observation and then look at the most likely causes so we can begin the troubleshooting process and repair the problem. An automated HPLC system's flow path typically consists of: The Solvent Pickup Filters (in the mobile phase reservoirs); The Pump(s); AutoSampler; AutoInjector; Column and one or more Detectors.*You should have a good understanding of this flow path before you proceed to diagnose the problem(s).

 *A gradual increase of pressure for the same method over time is often due to column fouling or a dirty inlet frit (e.g. PTFE frit). This article specifically focuses on the causes of a sudden change, not a slow change over time.

   Sudden System Back Pressure Changes: We will assume that you have been running the same method for some time or at least several times without a problem and then suddenly notice that the back pressure has changed from what is normally seen. The problem must lie within the flow path of the system.

   Excessive High Pressure: Typical reasons for this are:
  1.      A fouled or plugged column;
  2.      Wrong flow rate (higher than normal);
  3.      Inlet frit/filter plugged or restricted;
  4.      Plugged line;
  5.      Wrong mobile phase composition.

   Large Drop in Pressure: Typical reasons for this are:
  1.      A leak at a fitting, column or line (Number one reason);
  2.      Wrong flow rate (lower than normal);
  3.      Wrong mobile phase composition. 
  • Start by checking the method parameters to insure that they have not changed (i.e. flow rate, mobile phase composition). Check for leaks or plugs. If the column is suspect, replace it with a zero dead volume union (ZDU) and restrictor and flush the system. Replace the column with a new one or wash the current column according to the column manufacturer's guidelines.

   Sudden Peak Shape Changes: We will again assume that you have been running the same method for some time or at least several times without a problem and then suddenly notice that the peak shape of one or all of the peaks has changed from what is normally seen. *The key thing to keep in mind is that the change occurs all of a sudden, not because of poor initial method development.

   Typical reasons for this are:
  1.      Tailing or Split Peaks: Sample overload, change in flow rate, mobile phase composition (e.g. composition or pH), void formation, dirty frit, injection solvent too strong or a fouled column.
  2.      Fronting: Commonly seen when overloading sample on column.
  3.      Ghost Peaks: Usually due to a contaminated mobile phase, contaminated sample vial or contaminated injector (e.g. rotor seal).
  4.      Broad Peaks: Large sample injection volumes or extra column volume (bad connections with the system or tubing) are usually to blame. Try reducing the injection volume by a factor of 10 and see if the problem goes away. You may also want to wash the column as it may be fouled with sample.

   These are just two common problems we see when using HPLC systems. Note that a dirty or fouled column can cause many of these problems so take care of your columns and wash and test them regularly to insure they are in compliance. There are many other commonly seen problems besides these. If you would like to see a specific problem featured on this blog, then please send me a request.



  

Saturday, November 9, 2013

HPLC Maintenance & Repair Parts To Have on Hand for HPLC Systems

HPLC (UHPLC) systems are complex instruments which require periodic inspection, cleaning and maintenance. These tasks are critical to maintain the performance, reliability and accuracy of the instrument. If you have not done so already, I strongly recommend that you create formal standard operating procedures (SOP's) which address: (1) The frequency of when routine and non-routine maintenance procedures should be performed; (2) The types of maintenance and/or repair procedures used (e.g. piston seal replacement, A/I rotary valve seal replacement); (3) The exact step-by-step procedure to follow in performing these tasks and (4) The Performance Verification or Qualification steps and procedures which are to be performed to verify that any repairs made have been done correctly. *An instrument log book should be employed to document these procedures over time.

Periodic "General Maintenance" of the HPLC is one type of service procedure which should be scheduled at a set frequency (Example: Every 6 months) and will serve to provide a time to clean, inspect and repair/replace any parts which are worn due to normal use. Such routine HPLC maintenance is often referred to as a basic "Preventative Maintenance" service (or "PM Service"). Spare parts common to your HPLC system(s) should be on hand to perform these scheduled maintenance procedures as part of a normal PM service.

Here is a list of common parts that should be on hand for a "typical" HPLC system used in a pharmaceutical laboratory. Please consult the appropriate manufacture's product literature to determine the correct parts needed for your own HPLC system. This list is presented as a general guideline only:

  • Capillary tubing, fittings (nuts and ferrules): Assorted fittings, usually made of 316 Stainless Steel, but could be made of polymeric materials. Always have spare precut and polished chromatography tubing of appropriate I.D. and lengths for use with your HPLC available at all times. Insure that the nuts and ferrules used are appropriate for your brand of HPLC system and the columns used as different manufacturers have different specifications for their fittings and ferrules. Many types are not interchangeable.
  • Detector Lamps: At least one spare bulb of a type designed for your specific detector should be on hand. Note that some detectors use multiple lamps so you may need to have more than one type available for each detector. Some lamp bulb types (e.g. tungsten) can be safely stored and last for several years while other types, such as Deuterium bulbs, loose substantial energy after as little as 6 months. If you have several detectors of the exact same design, then there is often no need to stock multiple replacement bulbs for each one. Instead, stock enough bulbs to service one detector as it is unlikely you would see failure of more than one detector on the same day (an exception to this guideline is if you perform PM services on all of the instruments at the same time, then you may want to have multiple bulbs available).
  • Pump Pistons: One set of spare new pistons should be kept on hand for each pump module. As with lamp bulbs, if you have several identical pumps, then there is often no need to stock multiple sets of pistons for each one. Stock only as many as you expect to use in one year. Clean and inspect the pistons during each PM for any signs of scratches or surface abrasions. Under routine use, pistons should only require general cleaning and last a long time before replacement is required (> 1 year). Mobile phases which contain high concentrations of salt buffers often accelerate this wear requiring more frequent replacement. *Always install new piston seals when replacing pistons.
  • Pump Piston Seals: At least one set of spare new piston seals should be on hand for each pump module. Seals wear out more frequently than pistons. You should go through two or more sets of piston seals before you need to replace the pistons. If the piston seals leak, inspect the pistons for wear (replace with new ones or clean and reuse) and install new piston seals. Mobile phases which contain high concentrations of salt buffers often accelerate this wear.
  • Solvent Pickup Filters: These are the large particle filters which sit inside your solvent or mobile phase bottles. They are often made from stainless steel or sintered glass with porous inlets (~10 to 30 micron) and can clog or become fouled over time (esp. when used with aqueous buffers). In some cases these can be cleaned using sonication (not sintered glass filters, only steel or polymeric!). Note: Sometimes it is most cost effective to replace them with new filters then clean and re-use them.
  • Inline Frits/Filters: You may have an inline filter placed after your PUMP head, but before the column inlet to collect any remaining particulate matter. These filters can extend the lifetime of the entire HPLC system (esp. the A/S, A/I and Column), but will only do so if changed on a regular basis. Some manufacturers incorporate this type of filter into the design of their pump modules. An example of this can be found on the HP/Agilent brand model 1050, 1100 and/or 1200-series pumps. These have an inexpensive 10 micron PTFE frit installed in the outlet valve of the pump. This filter catches all of the normally occurring piston seal debris and larger mobile phase particles and should be changed every month. Other pre-filters are installed in cartridges just before the column inlet. These often overlooked pre-filters filters must be replaced about once each month to do their job properly. Keep plenty of spare filters on hand.
  • Auto-injector Rotary Valve Seals: If you have an auto-injector, then a high pressure valve is probably used to switch the sample into the flow path for analysis. This valve will have one or more parts which require regular inspection, cleaning and periodic replacement. Mobile phases which contain high concentrations of salt buffers often accelerate this wear. The valve rotor seal is the most common part which requires replacement.
  • Auto-Sampler Needle: A needle should last a very long time, but depending on the frequency of use and type of vial septa encountered it can require replacement at regular intervals. A good general guideline would be to keep one spare needle on hand for every 2-4 systems.
  • Auto-Sampler Needle Seat: The needle seat often requires more frequent replacement than the needle due to repeated mechanical wear. A good general guideline would be to keep one spare needle seat on hand for each system.
  • UV/VIS Detector Flow Cell: While not actually a required PM spare part, this one is worthwhile to have. If you employ a UV/VIS flow cell, then I always suggest you keep one dedicated spare flow cell on hand which matches the size and volume of the type you use in your instrument. A spare flow cell can prove to be very valuable as a troubleshooting tool if you believe that you have contaminated or clogged your current flow cell. A quick swap can answer the question and get you back to work quickly saving hours or days of lost time. *Note: This extra flow cell should be kept separate from all instruments for use as a tested spare only and not used for regular analysis.
If you have suggestions for other types of common HPLC spares to add to the list or to have on hand, then please let me know.

Saturday, October 12, 2013

What type of Water Should I use for HPLC, UHPLC or LC/MS Analysis?

Water is one of the most common solvents used in reversed phase chromatography. HPLC and LC/MS work demands ultra pure quality water be used in all applications which call for it as part of the method. Special types of HPLC analysis, such as amino acid analysis and ion chromatography, demand fresh ultra high quality water be used or artifact peaks may result. Poor quality or low grades of water may lead to "ghost peaks", baseline instability, high background noise or signals, contamination of columns and an inability to obtain reproducible results. Use the freshest and highest purity of water for best results.

A good starting point for describing the type of water suited to liquid chromatography applications is to look at the specification for ASTM Type 1 Reagent grade water. We often exceed this requirement for chromatography applications as several unspecified items such as nitrates and other chemicals present may have a negative effect on our analysis methods.

How does the grade of water affect our chromatography? The grade specified often dictates the amount of organics, bacteria, particulate, residues and overall absorbance the water will have. For example.

(1) Organics: High levels of T.O.C. can accumulate on the particles, inside the pores, or bind to active sites on the support inside the column causing a loss of resolution or sensitivity. *Lower T.O.C. levels are desirable.

(2) Bacteria: Microorganisms can contaminate the buffer solutions used causing ghost peaks, column fouling and the release of additional foreign organic matter into the system. This can result in clogs, ghost peaks, poor reproducibility or loss of resolution and/or sensitivity. *The water should be filtered through a 0.2 micron filter before use. Refrigerate solutions for no more than 3 days to slow growth, then dispose of the solutions.

(3) UV absorbance: High background or interfering ions which absorb can raise the baseline and noise levels seen, decreasing the total dynamic range. *Again, the lowest values, esp. at 200nm, are desirable.

A few of the general requirements for HPLC grade ultrapure Type 1 water can be stated as follows:

   Resistivity :         > 18 MΩ•cm at 25.0 C
   T.O.C. :              < 5 ppb
   UV cutoff :          190nm (as low in absorbance as possible!)
   Filtered :             0.2 micron Filter

*Some suppliers will also specify residue after evaporation (usually < 2 ppm); Trace metal analysis; Optical properties at specified wavelengths and other information. If purchasing by the bottle, request a copy of the lot certification sheet for the water so you can compare the measured values to other products.

Generating your own in-house, reverse osmosis (RO) ultra pure water from potable tap water is one of the best ways to insure you have high quality water for your LC methods. These systems pre-filter the water to remove large particulates then typically use UV lamps and/or multiple resin cartridges to remove the maximum amount of T.O.C.'s from the water plus many trace metals before finally filtering the water through a 0.2 micron membrane as a final polishing step. Various types of systems can be purchased, but for HPLC or LC-MS applications, it is critical that you select a system that provides ultra pure water suitable for your applications. Periodic maintenance of the filter cartridges and monitoring of the main water supply source is critical to their operation (some "tap" water sources may require pre-treatement). *"Water On Demand" systems such as these provide fresh clean water on demand so there is no need to be concerned with storage issues. A number of different vendors offer these lab grade systems for HPLC and LC/MS applications and you can contact them (e.g. Millipore/Sigma Milli-Q® brand) to determine which system will provide you with the volume and quality of water which is appropriate for your application(s).

If you do not have access to an in-house reverse osmosis system, then purchasing HPLC or LC/MS grade water in glass bottles may be another option. A hint, before opening and using them,  clean the outside of bottles of all dust. Date the bottles when you first open them. Bacteria will start to grow once the bottle has been opened. The glass will also slowly leach ions (i.e. Sodium) over time into the water so it is best to use the water quickly.

Never underestimate how the quality of the water you use to perform chromatography can change the results seen in your methods. Water quality is just as critical as any other component in your system so be sure and take the time to monitor it just like you do to any other part of the system.