One of the more common LC/MS problems I am asked to help solve deals with contaminated LC-MS or LC/MS/MS systems. Over time, many systems will become contaminated with a wide variety of plasticizers, detergents, salts, metals and ion pairing agents that routine source cleaning will not remove. Often, these compounds are introduced to the system through the tools used (e.g. pipettes) chemicals, solvents, mobile phase additives or even the samples themselves. "Dirty" samples sometimes persist inside the system long after the analysis work is complete, leaving material in poorly maintained injection valves but also through the use of poorly washed / contaminated and fouled HPLC columns. Even the modern inline HPLC vacuum degasser has proven to be a source of contamination.
In addition to the above mentioned sources of contamination, another more obvious source of contamination should always be addressed early in the process of cleaning the system. Specifically, the glass mobile phase bottles and the associated solvent pickup tubing and solvent pickup filters used with them. Contamination in these areas may directly infuse the system with undesirable material. Good cleaning and maintenance practices must be maintained to reduce this source of potential contamination.
As a general guideline, we shall not place our mobile phase reservoir bottles in any type of dishwasher or wash them using any dish soaps. These may leave a residue easily detected by even the weakest mass spectrometer. Avoid contamination by purchasing high quality glass bottles with vented caps to keep dust out. If rinsing with organic solvents (and/or freshly prepared and filtered high resistance water) does not clean them, you can try a Nitric Acid rinse (up to 30%) followed by a neutralizing wash in 2M Sodium hydroxide. Follow-up with many rinses of HPLC Grade water (or LC/MS grade), oven drying, then re-fill with an appropriate mobile phase. Don't forget to replace those solvent pickup filters too. While many 316 SS pickup filters can be cleaned, most of the sintered glass style filters are designed to be disposed of (not cleaned or put in an ultrasonic cleaner!). So periodically dispose of the glass types and install new filters and fresh mobile phase into those recently cleaned bottles (before you start looking for the source of contamination in the more expensive parts of the instrument, clean or replace the filters). - Please don't re-contaminate an expensive HPLC or LC/MS system and invalidate your methods and data because you skipped replacing a $10 part. Keep commonly used spare parts in-stock and always maintain a clean system.
The HPLC "Hints and Tips" found here will be of interest to anyone performing HPLC, UHPLC (UPLC), LC/MS, SFC, CE or GC analysis. Many of these free tips, taken from professional classes that I teach, discuss fundamental to advanced level chromatography concepts in a practical way. Occasionally, I will also include laboratory computer related tips here too. Please "follow" this blog to receive updates of new posts. Hint: Type keywords into the SEARCH bar to quickly find other articles.
Translator for HPLC HINTS and TIPS for Chromatographers
Saturday, April 8, 2017
LC-MS Contamination? Another Possible Cause. Are your Mobile Phase Bottles and Pick up Filters Clean ?
Saturday, March 4, 2017
The Three Most Common HPLC Questions and How To Solve Them
The three most common HPLC related questions I am asked each week can be summarized below. Test your basic chromatography knowledge. Before reading the answers, see if you can answer them correctly on your own.
- "What Is Causing the HPLC Baseline, Pressure or Peak Retention Time(s) To: Wander, Change, Drift, Vary or be Unstable?"
- "How Should I Wash or Regenerate My HPLC Column?"
- "How Can I Tell if the Sample is Retained On the HPLC Column? or What Does It Mean When No Chromatography Took Place?"
Let us address each question in order and attempt to provide accurate answers (I have included links after each question to articles with more detailed explanations).
What Is Causing the HPLC Baseline, Pressure or Peak Retention Time(s) To: Wander, Change, Drift, Vary or be Unstable?
- Retention times must be reproducible from run to run.The causes of an unstable baseline and/or changing peak retention time(s) are often related. Common reasons include: Column temperature fluctuations, inadequate mobile phase mixing or degassing, leaks, dirty column, sample overload, lack of pH or buffering control (weakly ionizable samples can be very sensitive to changes). *Full Article link with detailed answers, here.
How Should I Wash or Regenerate My HPLC Column?
Note: Before proceeding with any column regeneration or cleaning procedures, always refer to the specific advice provided by the column manufacturer. Approved maintenance and cleaning instructions can often be found in the product guide or booklet which comes with the new column. Additional information can be found on the vendor's website or by contacting them directly.
- Two issues must be addressed to answer these types of questions. (1) Always wash your column with a specific column wash solution which is stronger than your analysis solution. The use of a stronger solution (In this context, "stronger" means better at dissolving the samples and faster at eluting them from the column) as the wash solution requires regular use to maintain the column. Failure to regularly wash your column may result in compounds accumulating on the column over time (fouling the column) resulting in poor reproducibility, higher back-pressures, contamination and/or poor peak shape. (2) Next, always wash your column after each analysis. This should be a separate step, not incorporated into your analysis method. The analysis method should not include the column re-equilibration steps at all. A second, separate wash method should always follow each analysis method which includes the rinsing of the column with a "stronger" solution for an adequate period of time, then adjustment back to initial conditions where re-equilibration can take place to get it ready for the next analysis run. These are fundamental guidelines of good method development and follow well established principles. Developing methods in this way should increase the lifetime of your columns and improve the reproducibility of results obtained (better %RSD run-to-run).
For more information on washing bound proteins off RP HPLC columns, please refer to this linked article found here.
How Can I Tell if the Sample Is Retained On the HPLC Column? or What Does It Mean When the Sample Comes Out At or Near the Column Void Volume?
- Chromatography is a tool which when used properly adds one or more additional dimensions of physical or chemical characterization information to your analysis data. It does so first by using on-column RETENTION. Samples must be run under conditions which allow the material to interact with the chromatography support for a period of time. We define this time as the retention time. A sample which does not interact at all with the column support material will elute off the column early (and not be retained) at the "column void time" (or column dead time). We refer to this void time as the "T zero" time. When a sample elutes at or near the T zero time, no chromatography has taken place and no method has been developed. It is as if the HPLC column was not used. How do you know what the "T zero" time is (it will be different for different methods)? You must first calculate the HPLC column's dead volume. Once you know the column dead volume and flow rate, you can calculate the T zero time. A scientifically valid HPLC method will include conditions which retain the sample on the column for a long enough period of time to insure that it is interacting with the support. This allows for separation from other compounds to take place and is the purpose of chromatographic resolution. Without this retention mechanism, you are just flow-injecting the sample past the column and skipping all chromatography. It would be far simpler to just place the sample in a spectrophotomer cell as no retention or additional data would be obtained using that technique.
- When first learning liquid chromatography, two of the very first calculations you must learn to use in HPLC are: Column Dead Volume (aka: Column Void Volume) and the K prime of a sample (aka: Peak Capacity Factor). Do you know how to calculate these? They are calculated and reported for each method used. You should be able to tell anyone who asks you what the values are for each method. A chromatographer must know and understand them before using an HPLC system or running a method. They are also critical to method specificity and proper validation. Here are links which after reading and practicing, should make you an expert in these two fundamental calculations.
So, how did you do answering these basic questions? If you have put in the needed study time and practical experience to learn and use these fundamentals of high-performance liquid chromatography, then you should have been able to easily provide correct answers to all three questions. If not, then it is time to go back and study up on those basic liquid chromatography texts and article links, plus get more supervised hands-on time with the instruments.
Labels:
Baseline unstable,
Column,
Dirty,
Drift,
Fix,
Fluctuation,
Fouling,
HPLC,
K Prime,
Pressure,
Problem,
Problems,
Question,
Questions,
Retention,
Solve,
Troubleshoot,
Troubleshooting,
UHPLC,
UPLC
Saturday, February 4, 2017
Determine the HPLC System Dwell Volume (Gradient Delay Volume)
Note: The total HPLC gradient system
dwell volume is different than the HPLC column’s void volume. Two different terms for two very different measurements.
When we perform gradient HPLC analysis, the mobile phase composition is changed over a period of time. The mobile phase is mixed in real time by the pump(s), mixer and/or valves, then transported to the injector and finally, on to the head of the HPLC column. The total volume of liquid contained between where the mobile phase is mixed and the head of the column helps us determine when the newly mixed solution arrives at the column head (it is not instantaneous). This delay is often referred to as the gradient delay time (or delay volume) and its value will vary for different HPLC systems due mainly to differences in tubing dimensions used, pumping system type and the design of the flow path.
When we perform gradient HPLC analysis, the mobile phase composition is changed over a period of time. The mobile phase is mixed in real time by the pump(s), mixer and/or valves, then transported to the injector and finally, on to the head of the HPLC column. The total volume of liquid contained between where the mobile phase is mixed and the head of the column helps us determine when the newly mixed solution arrives at the column head (it is not instantaneous). This delay is often referred to as the gradient delay time (or delay volume) and its value will vary for different HPLC systems due mainly to differences in tubing dimensions used, pumping system type and the design of the flow path.
For
example: If the system dwell
volume is found to be 1 ml and the flow rate used is 1.000 ml/min, then the
gradient delay time is one minute.
So how do we
know what the system dwell volume or gradient delay volume is? Well, we measure
it of course!
Measure the ‘System Dwell Volume’ (aka: Gradient
Delay Volume)*:
(1) REMOVE any
HPLC column(s) and install a Zero Dead Volume Union (*ZDV) or a restriction capillary of know volume in its place.
(2) Prepare Two
Different Mobile phase solutions:
Bottle
‘A’: HPLC grade Methanol (MeOH).
Bottle
‘B’: HPLC grade Methanol with 0.1% acetone added (v/v).
(3) Set your
UV/VIS detector to 265 nm (8 nm Bandwidth, Reference OFF).
(4) Program a
suitable system flow rate and create a simple Gradient Method (linear change)
which starts at 0.0 minutes with 100% ‘A’ (HPLC grade Methanol) and 0% B (HPLC
grade Methanol with 0.1% acetone added) and runs to 0% ‘A’ and 100% ‘B’ for about
10.0 minutes (actual times used will depend on your selected flow rate).
(5) Flush and degas
both solutions, ‘B’ first, then ‘A’ through the system until you get a nice
clean, flat baseline. Make sure their is enough backpressure on the pump (>40 bars) to obtain a stable signal (use a restrictor or back-pressure regulator if needed).
(6) No
injection should occur during this method.
(7) Start the
method (RUN) and observe the 265 nm signal over time. At some point you should
observe the signal begin to rise. When you see this signal change occur, the
acetone has finally made it from the pump head to the detector’s flow cell.
Make note of the time this occurs.
Using the known flow
rate and observed signal change time, you can now estimate the total system dwell volume.
Example:
If you observe the signal start to rise steeply at 2.00 minutes and your flow
rate was 1.000 ml/min. Your system dwell volume would be 2.000 mls.
A more accurate
system dwell volume value can be obtained by next running the same method with an
injection of acetone (e.g. 1 ul) and noting the time at which the injection
peak is first seen. That will give you the time it takes the sample (and therefore
the volume needed) to go from the injector to the flow cell. If you subtract this time
off the system dwell time you recorded in the last test, you will have the
actual measured time from the pump head (or proportioning valve) to the head of
the column (vs the flow cell). Normally the volume contained in this tubing and
flow cell are very small relative to the volume in the rest of the system, so
we can ignore them. However, when using some of the very low volume columns
(e.g. 2.1 x 50 mm), the volume contained in these areas can become significant
so when appropriate, we need to be aware of them.
Failure to take
into account changes in HPLC system dwell volumes can result in methods which
no longer work or provide different results. This is because the gradient
rate change you program in your method may not allow enough time for the new mobile
phase composition to reach and flow all the way through the column in the time
that you have programmed. A common mistake we see is when users forget to
adjust the gradient profile when changing column dimensions or program changes
using too fast a time.
BTW: One common
trick we use to improve compatibility between systems which have different
dwell volumes is to include an initial (time 0.0) isocratic hold-time into the start of each
method. If all systems used have system delay volumes under 3 mls, then add a 3
minute isocratic hold time at the start of each method (if 1.000 ml/min flow
rates are used), before any gradient starts. While not the best way to deal
with the issue, this type of “cheat” can make it possible to quickly adapt a
method for use on several different system types.
*Note: This is a generic method to determine
the system dwell volume or gradient delay volume. Detector signal buffering and flow cell volume also adds to the delay and in some cases, must also be accounted for too. There are many other methods
which can be used for this determination as well. This proposed example serves to illustrate the
concept only.
Labels:
Chromatography,
Dead Volume,
Delay,
Determination,
Dimensions,
Dwell,
Elute,
Gradient,
Gradient Delay,
HPLC,
HPLC Column,
Resolution,
Retention,
System,
T zero,
Time,
UHPLC,
Void,
Volume
Saturday, December 31, 2016
PEAK PURITY Determination by HPLC Diode Array Chromatography Software (UV/VIS): Limitations and Uses
"Peak Purity"
software determination by HPLC UV/VIS detection is one of the most abused and easily misunderstood
features found in advanced liquid chromatography systems (e.g. HPLC, UHPLC and
CE).
For HPLC, one or more inline detectors can be used which provide additional
data about a fully resolved peak’s physical or chemical properties. The data obtained can be
compared to that of a pure standard, or known impurity. For compounds which absorb light in the region of most
UV/VIS detectors (~ 200 to 900 nm), a single
wavelength detector (e.g. UV/VIS) provides a very limited
second dimension of data (retention time is the first dimension), but a scanning, multi-wavelength UV/VIS detector can
add a second and third dimension of data to the retention time. Scanning
detectors, commonly known as Diode-Array Detectors (aka: DAD or PDA) are
commonly used in HPLC and CE analysis (they are required for routine method
development). A scanning DAD can provide detailed sample UV/VIS
spectra across a range of wavelengths for each peak, at any retention
time recorded, allowing for a 3D plot of the spectra to be recorded much like a
“fingerprint”. "Pure" compounds which do absorb light across a pre-defined
wavelength range should show identical spectral profiles (“slices”) across the
upslope, apex and down slope of the resolved peak. "Impure" peaks may
show dissimilar spectra across the width of the peak revealing the presence of
a co-eluting peak or impurity. Impure peaks may also NOT show any dissimilar spectra at all (because some compounds may not be detected). When a properly developed HPLC analysis method is used to evaluate the purity of a sample, the
single dimension of “retention time” is evaluated with additional dimensions of analysis such as the UV/VIS peak spectra. "Peak Purity" relies on the detection of a sample's spectral profile to detect the presence of an "impurity" (that may have co-eluted with the sample). This additional
dimension of analysis (full Spectra) is required to improve the confidence level that a peak may in fact be correctly identified (qualitatively) and does not contain any co-eluting compounds. IOW: "Peak Purity" does not actually test for purity.
Diode-Array 'Software' based Peak purity
determination by HPLC is a qualitative assessment of the impurity profile of
the sample. It is designed to reveal impurities, NOT prove peak purity.
BTW: We really should rename it “Peak Spectral Impurity Assessment"
because that is in fact what we are measuring. The algorithm used for Peak
Purity determination is designed to confirm the presence of one or more
impurities by comparing spectral data slices (multiple slices taken at the apex
and both the upslope and down slope sections of the peak). A mismatch
would indicate the peak has not been fully resolved (one or more co-eluting
peaks are present). In other words, it is impure by UV/VIS
analysis. Note: It does not indicate that the compound is impure, but rather
'the peak' being measured is. As you can see, the concept makes sense, but the how it is used in many laboratories is flawed leading to invalid reports and data.
- “Peak Purity” does not in fact indicate the actual purity of the compound, but instead indicates when a peak may be found to contain impurities. It is an estimated measure of PEAK Impurity.
In simple terms, IF the spectral
slices obtained from one peak are not identical, than the peak may
contain one or more impurities. Co-elution is the most likely reason for this.
Points to consider when using
"Peak Purity" software:
- The absence of any spectral differences across the sample peak are not an indication of actual purity;
- Compounds similar to your sample may have similar absorbance profiles (fooling the system);
- The relative concentration of actual impurities may not be high enough to detect;
- The compounds / impurities may not absorb light at the wavelengths scanned;
- The HPLC
method used, the software settings and the parameters that you chose in
the ‘Peak Purity’ software menu have a huge effect on the results obtained.
Different people often get different results for the same sample. Inputting poor quality settings or using a poor quality method often leads
to misleading purity results. This is an advanced software feature requiring many years of training to use. Again, it does NOT test for purity.
- The peak of interest must be retained on the column (K prime > 2) and resolved apart from any observed peaks. Don't use peak purity to analyze peak(s) which elute at or near the column void volume (Low K prime values may demonstrate that good chromatography fundamentals were ignored. Poor quality methods fail validation). Poor quality HPLC method and poorly selected DAD "Purity" settings result in invalid results (audits, recalls etc may result from reliance on a subjective "software" feature).
We prefer to think of HPLC 'Peak
Purity Assessment' as a null test. If the recorded peak spectral data slices are
different, than you probably have co-elution and/or impurities present (so try
and develop a better method to resolve the peaks apart). If no differences in
the spectra are seen (they are similar), then the peak may be pure or may
contain compounds with similar spectra as are commonly seen with related
reaction synthesis products or compounds. So only when you detect differences
in the acquired spectra can you be confident that there IS a qualitative
difference or impurity present. You will not know what percentage of impurity level is (since you do not know what it is).
When configuring the Peak Purity parameters for your sample, you must start with a very high quality HPLC method (A "validated method" is not necessarily a high quality method. "Validation" does not in fact insure that the method follows good chromatography fundamantals). The correct detector sample rate, threshold, slope, signal wavelength and bandwidths need to have been properly selected and used (Reference Wavelength always OFF). The peaks shown in your chromatogram should have excellent symmetry with good on-column retention (K-prime, as applicable to mode), baseline separation (> 2.0 for non-SEC modes) and very low baseline noise levels. The two Peak Purity spectral reference points should be manually selected and placed at times before and after the peak of interest in clear baseline areas where no other peaks or spectra are seen (never use the instrument default settings for reference points!). Select at least 7 spectra from the sample peak for comparison (more detail can be provided with more spectra, but be careful not to select spectra near the baseline or the noise limits). If your method and chromatogram are not of the highest quality, then please do not use the automated "peak purity" analysis feature, instead spend time improving your method.
When configuring the Peak Purity parameters for your sample, you must start with a very high quality HPLC method (A "validated method" is not necessarily a high quality method. "Validation" does not in fact insure that the method follows good chromatography fundamantals). The correct detector sample rate, threshold, slope, signal wavelength and bandwidths need to have been properly selected and used (Reference Wavelength always OFF). The peaks shown in your chromatogram should have excellent symmetry with good on-column retention (K-prime, as applicable to mode), baseline separation (> 2.0 for non-SEC modes) and very low baseline noise levels. The two Peak Purity spectral reference points should be manually selected and placed at times before and after the peak of interest in clear baseline areas where no other peaks or spectra are seen (never use the instrument default settings for reference points!). Select at least 7 spectra from the sample peak for comparison (more detail can be provided with more spectra, but be careful not to select spectra near the baseline or the noise limits). If your method and chromatogram are not of the highest quality, then please do not use the automated "peak purity" analysis feature, instead spend time improving your method.
SUMMARY:
The HPLC UV/VIS
Peak Purity Analysis (“Peak Spectral
Purity”) feature is very complex and has many software settings which must
be set up correctly to obtain any scientifically useful data regarding possible peak impurity levels.
- Do NOT use the system default settings / values for 'Peak Purity' ! They are just place holders for actual values (which you must calculate and fill in the correct values for your method).
* Due to a general lack of formal training, I often see this software feature being used incorrectly by
most chromatographers. This is worth repeating... the HPLC method
used to obtain the original data must be of the highest quality and the
training of the operator must also be at the highest level. To use this advanced software feature
successfully, an advanced understanding of the fundamentals of chromatography
are required as are a detailed understanding of all of the peak purity software
features (how to set the correct threshold, obtain reference baselines, Set
sampling rate, noise levels, signal extraction, normalization settings…).
Routine HPLC training classes do not cover these types of tasks. Years of specialized
training and practical experience are required to use these tools. Never use
the “automated” versions or the manufacturer’s default values to find “Peak
Purity”. The only correct way to use these features is to manually tune the
method and settings to your specific sample. Failure to customize the method
and settings used may result in invalid data and incorrect "purity"
determinations.
Due to very complex software setup needed for "Peak Purity" determination by UV/VIS spectra, the requirement for a high quality HPLC method and a high quality data-set,it is our opinion that few should ever use it. In general, the recommendation for most chromatographers is to
not use this feature unless first having
demonstrated the required skills and advanced understanding of the fundamentals
of chromatography. Most of the methods that we professionally review where "Peak Purity" data have been used as part of the method have been found to be based on invalid methods, resulting in any "purity statements" issued as unscientific and invalid. Please proceed cautiously and request professional review of any methods which employ it BEFORE committing to relying on it.
©Copyright, March 1, 1996 by
William Letter of Chiralizer Services (Plainsboro, NJ) from a portion of
material presented in an HPLC Diode Array Method Development Class.
Labels:
CE,
DAD,
Determination,
Determine,
Determining,
Diode Array,
HPLC,
Impurity,
Limitations,
PDA,
Peak,
Peak Purity,
Purity,
Software,
Spectral Purity,
UHPLC,
Use,
UV
Subscribe to:
Posts (Atom)