Saturday, December 26, 2015

Common pKa Values for ACIDS & BASES used in HPLC and LC/MS Method Development





pKa (25°C)                              ACID
0.3                                           Trifluoroacetic acid
2.15                                          Phosphoric acid (pK#1)
3.13                                          Citric acid (pK#1)
3.75                                          Formic acid
4.76                                          Acetic acid
4.76                                          Citric acid (pK#2)
4.86                                          Propionic acid
6.35                                          Carbonic acid (pK#1)
6.40                                          Citric acid (pK#3)
7.20                                          Phosphoric acid (pK#2)
8.06                                          Tris
9.23                                          Boric acid
9.25                                          Ammonia
9.78                                          Glycine (pK#2)
10.33                                        Carbonic acid (pK#2)
10.72                                        Triethylamine
11.27                                        Pyrrolidine
12.33                                        Phosphoric acid (pK#3)



Notes: (1) This is a general list of commonly used acids & bases for chromatography applications and not meant to be a comprehensive list of all values. (2) TFA is an overused and very strong acid for many chromatography applications. It also has strong ion pairing properties and can result in high UV noise, vacuum degasser and/or MS contamination. If you must use it, try and use the lowest concentration which results in the desired pH. Example: 0.1 % TFA ~ pH 2.0, 0.02% TFA ~ pH 2.7. (3) Formic acid is a popular alternative to TFA for many applications, esp LC/MS. (4) Not all acids/bases provide "buffering" on their own.

Reference: CRC Handbook of Chemistry & Physics.

Tuesday, February 12, 2013

pH Measurement of HPLC Mobile Phase Solutions and Buffers

Several times each month I am asked how to "correctly check and adjust the pH of an HPLC buffer solution which has an organic solvent component"? Well, the answer is to always check and adjust the pH of the purely aqueous solution first. Only pure aqueous solutions can be correctly adjusted for pH in the laboratory. Do not mix any organic solvent into the water based solution until after you have correctly adjusted the pH. The addition of an organic solution will throw off the final reading. Once the aqueous portion of your solution has been correctly adjusted to the desired pH value, then you can mix the solutions (or run an organic solvent gradient against the aqueous portion) as needed.

*This procedure also serves to make sure that all solutions used in chromatography are prepared in the same manner. It is true that the pH of the final mobile phase mixture (aqueous and organic mixture) may not be the same anymore, but the prepared stock solutions from which they were made will be the same each time, insuring reproducible results. Developing and describing chromatography methods and procedures which are highly reproducible equates to good scientific technique.