Saturday, April 9, 2022

Agilent Quaternary Pump (e.g. G1311A ) "Secret" Operator Tip to FLUSH the HPLC Pump in 1/2 the time!

One of the most popular "tips" taught in our Agilent 1100 and 1200-series HPLC training classes shows users how to speed up the daily priming and flushing process of the Quaternary Pump. Many people use these pumps without taking advantage of the Quaternary pump's higher flow capability. If you are not currently using the higher 10mL/min flow rate capability offered by this pump (vs. the Binary pump's 5 mL/min), then you are missing out on a free time saving feature. Please read on to learn how to use this feature.

Based on the HP 1050 pump and introduced in 1995 as the "1100-series" version, the G1311  "Quat" pumps are one of the most popular research grade HPLC pumps found in laboratories today. They are extremely reliable, rugged, easy to operate and service. The Quat pump is driven by an easily accessible, single pump head with an in-series, servo controlled dual plunger and Multi-channel Gradient Valve ('MCGV') for 4-channel solvent proportioning with an active inlet valve (known as the 'AIV', first used in the HP 1050 pump and the reason for this pump's high reliability. No more "sticking" inlet valve issues!). Unlike the Agilent Binary pump (G1312), which uses two separate dual plunger pumps (2-channel) at up to 5.0 mL/min (maximum), the Quat pump offers an extended flow range, up to 10.0 mL/min (maximum). However, most users are not aware of this or do not know how to utilize this higher flow rate feature because the Quat pump defaults to a maximum flow rate of 5 mL/min at initialization. The ability to program the pump to operate at flow rates greater than 5 mL/min requires a "trick" to activate it (which apparently is a secret as we rarely encounter customers who are aware of how to use it). 

Let me share with you why you would want to use this feature, why the feature is hidden to most and of course HOW TO ACTIVATE IT on the Quat pump.

  • Q: Why would you want to run the pump at 5 to 10 mL/min? Semi-prep columns can be run within this flow rate range, but a more common reason to operate at 10 mL/min is for daily system start-up. Anytime you replace or change the mobile phase bottle/solution OR when you startup the HPLC system (each day) one of the very first things you need to do is prime or flush each of the mobile phase channels, one-at-a-time through the system to waste. Air bleeds into the system when it is not used and this procedure primes the lines and pump head with fresh mobile phase preparing it for use. The system's flow path is directed to waste (via the open, prime-purge valve) during this step so back-pressure is not a concern. The higher the flow rate you can use for this flushing step, the sooner you can complete it. If you run the pump at 10 mL/min vs 5 mL/min, then flushing can be completed in half the time. This is especially useful if you have a model G1322A degasser module installed as the internal volume of each degassing channel in the G1322A is 10-12 mLs, requiring extended flushing times (4x channels = 30+ mLs flush per channel) before moving on to the next channel.
  • Q: Why does the Quat pump initialize with a reduced, 5 mL/min maximum flow rate? The Quat pump was designed to meet two different operating pressure ranges. From 0 to 5 mL/min the permitted operating pressure range is 0 - 40 MPa (0 - 400 bar). Above 5 mL/min, the operating pressure range is reduced, 0 - 20 MPa (0 - 200 bar). As most analytical chromatography is performed at flow rates below 5 mL/min, the system initializes using the more practical, 0 - 400 bar range, limiting flow rates to 5 mL/min maximum. The default maximum pressure field is set to 400 bars. You should always change the maximum pressure value from 400 bars to a more realistic maximum pressure (lower value) for your method. Use a maximum value that is appropriate for your own method. *The only time you will want to set it to the maximum value is when conducting a Pump Pressure/Leak test (it must be set to max pressure for testing).
  • Q: When I try and enter a pump flow rate larger than 5 mL/min, the system does not accept it. How do I program the pump to increase the flow rate past 5 mL to 10 mL/min? In order for the system to accept a flow rate of greater than 5 mL/min, you must FIRST set the maximum pressure limit to a value that is 200 bars or less (within the allowed "0 - 20 MPa (0 - 200 bar)" range). Once the maximum pressure limit has been reduced in the method, the system will then allow you to enter a higher flow rate such as 9.999 mL/min (10 mL/min). As long as the maximum pressure alarm is set within this window (200 or less), the pump will allow flow rates above 5 mL/min to be used. Now you can program the pump to flush lines or prime the system at twice the speed of the Binary pump equipped systems (10 mL/min).

Please share this "trick" with other users of the G1311A, G1311B, G1311C versions of this pump so they can maximize their time and productivity. Let us know if you find this tip useful.


Saturday, October 29, 2016

Notes on Cleaning bound Protein from RP HPLC columns:



First, a few comments:

  • ·         Before proceeding with any column regeneration or cleaning procedures, always refer to the specific advice provided by the column manufacturer. Approved maintenance and cleaning instructions can often be found in the product guide which comes with the new column. Their guidelines supersede these!
  • ·         Columns are consumable items. After a suitable amount of use, the time and materials required to regenerate them may cost more than the purchase of a replacement column. Always have a new, spare column on hand.
  •        Do not overload the column! This is the most common reason for column fouling, flow path contamination and sample carryover issues. In most cases, injection volume should be less than 1% of the column volume (maximum).
  • ·         Protect your detector. Before washing or cleaning the column, disconnect the column outlet line and direct the column to waste only.
  •        Column Storage solutions are not the same as column wash solutions. Never store a column in buffer or ion pairing containing solutions.

For RP supports, if buffers have been used, always start by washing the column down with ultra-high purity water and some organic solvent (e.g. Water/MeOH, 95%/5%) to remove all salts. Use about 10 column volumes to flush these off. Do not wash the column with organic solvents until you have first washed it thoroughly with high-purity filtered water.


Polymeric resins (e.g. PS-DVB) from many manufacturers can effectively be cleaned using 0.1 M Sodium Hydroxide solution or a mobile phase solution containing equal parts of isopropanol (IPA) and 1 to 3 M Guanidine hydrochloride at ~ 50 °C. Optionally, some success has been reported using other solutions such as: 5M Urea (pH 7) buffer solution; 1 M NaCl (pH 7) and even mixtures containing some methylene chloride solvent. Check with the manufacturer first as column damage/plugging may result if their directions are not followed.!

For RP silica based supports (non-SEC), we often start with a series of wash solutions. In most cases, pure water or pure organic solvents such as MeOH or ACN will not remove bound protein (common novice mistakes). An acid, base or even an ion pairing reagent is often needed to clean them. Start simple and monitor.
 
For RP silica based supports (SEC), a high salt buffer solution often releases bound proteins quickly. A mobile phase containing water plus an alcohol (methanol, IPA or ethanol) may also prove effective too.  Optionally, a solution of 0.5 M guanidine hydrochloride may effectively remove bound material.

General Advice: One of the first general wash solutions to start with (especially to remove basic compounds) is a 1% Acetic acid solution in Methanol (50/50). If desired a stronger acid such as 0.1 % Trifluoroacetic acid (TFA) or 0.1 % Formic Acid can be swapped for the acetic acid (where possible, start with a weaker acid). Use a low concentration of acid to achieve a pH of ~ 2.5. This acidic wash can be followed with a neutral solution, or if needed, a later solution where IPA or ACN replaces the MeOH used (50/50).

For extreme cases where the column has been overloaded with protein, a 5 M Urea solution has been proven effective in removing bound protein from silica and polymeric supports too. A word of caution, as the resulting pH of this strong solution may be greater than or equal to pH 9. Many types of silica based RP columns can not withstand strongly basic solutions and the silica inside may dissolve (plugging the column). Start with a lower concentration wash  first. You can always increase it later. Always read the instruction sheet carefully which came with the specific HPLC column to determine if it can be used at these high pH levels. Another salt solution that has shown some promise is 1 M sodium phosphate solution, pH 7.0. Run the salt solutions for about one hour at a moderate flow rate. Follow up all washes with rinses of mixtures of water and MeOH (80/20), then 90% MeOH/Water. 

Please remember that in ALL cases, HPLC columns are consumable items with a limited lifetime. Dispose of them properly when they are damaged or contaminated and replace with a new column. Once you have a fresh clean column to work with, prevent column fouling by developing better quality methods which utilize frequent, properly developed wash methods (using a wash solution which is stronger than your analysis mobile phase), filter all samples and be sure they fully dissolve in solution (100%). *Column fouling is not normal and can be prevented with proper training.

Saturday, August 27, 2016

Modern HPLC Method Development Tips (PART II):

This is the second of two articles (Part I) which will provide suggestions on how to improve the HPLC (UHPLC) method development process. - PART II



INITIAL METHOD DEVELOPMENT:
  1. Before you start, learn what you can about your sample, its hazards, solubility and properties. Conduct a quick literature and/or keyword search on the web using a popular search engine (e.g. GOOGLE). You can often find many journal articles, white papers, application notes and chemical data on the web in just a few minutes.
  2. Determine which liquids your compound(s) are soluble in. Use a pipette and several glass vials or tubes with different solutions (pH is important too). This will narrow down the types of mobile phase and chromatography modes that you can use.
  3. Column choice is the most important part of method development. The stationary phase that you choose has the single greatest effect on selectivity! Select the right column and mode of chromatography. Most RP methods should start with at least a modern ultra high purity, metal free type B silica column with a C8 or C18 support (*But you must select the column type that best suits your application). For small molecules (<1,000 Daltons) select a standard sized analytical column with a 2.5 to 5.0 micron particle size and pore size between 60 and 120 Ã… (e.g. 4.6 x 150 mm; 3.0 x 100 mm). *For larger peptide or protein molecules you will need a particle with a large pore size (~ 300 Ã…). For optimal results, columns with very small internal volumes should be paired with HPLC systems with similar ultra-low internal delay volumes. Your HPLC system should be optimized to balance the needs of good mixing, low delay volume and proper sampling rate for your method. If the method is likely to be transferred to different types of HPLC systems, then you may want to initially stay away from the < 2.1 u particle supports for your method development. At this time, these tiny particles can not be reliably packed into columns as well as the larger sized particles and generally have much poorer %RSD values than larger particles (i.e. 2.5 to 5 u). Method development is initially easier & generally more rugged using conventional particle sizes [Keep things simple when starting out. You can always change later on]. Once you have selected a column and decided on a flow rate, make sure you calculate and measure the actual column void volume to find the 'Tzero value' (column dwell volume). You will need this value to find out if your compound has been retained on the column (K prime) and to also determine most of the needed performance calculations and parameters.
  4. Once you have selected an HPLC column (if possible, please start with a brand new column) you will need to test it and establish a baseline to show that it meets both the manufacturer’s specifications and, far more importantly, that it meets your method’s requirements. In most cases, do not use the manufacturer’s QC test solution for this. Those stds are designed to allow the column to easily pass manufacturing QC, not the more critical requirements that you may need to prove. We prefer to use a real sample mixture (~2 compounds) that are similar to the type proposed for the method. They must be well retained on the column (K prime of > 2), easy to prepare, stable and reliable. This std and the specific method you develop for it, will be used to prove the column’s performance (i.e. R, S, K prime, Tzero, Plates). It will be used again, when the column’s performance is called into question.  
  5. Notes on Mobile phase and additives: Keep it simple. Avoid the use of any additives such as ion-pairing reagents when first starting out. They are overused in general and can cause problems later on. If required, they can always be added later on. Use only HPLC grade solvents, fresh RO HPLC grade Water and the highest purity acids, bases and/or additives, if required. Use only filtered (0.22u) products. Always dissolve samples in the mobile phase (or a weaker solution) for injection. For many RP methods a low pH mobile phase (~ pH 2.5) provides a good starting place. Samples with a pKa far enough away from this value are likely to be retained, stable and unaffected by small changes in pH. *pH is usually one of the last parameters which is optimized.
  6. Use a Gradient Method to find the approximate elution conditions of the mixture AND make sure you are detecting all of the compounds injected onto the column [For dedicated RP isocratic methods, start with a high organic mobile phase % (i.e. 95%) and record the results. Continue to reduce the organic content in steps of 10% and observe where the best compromise exists between retention and elution]. For RP gradient methods, start with a very high aqueous % (e.g. 98%) and run your gradient, slowly, to a very high organic % (e.g. 95% or 98%, not 100%!) to make sure you retain, hold, and then elute everything. Your mobile phase conditions must be strong enough to elute everything off the column during the gradient portion of the run (long enough hold time). Please Do NOT include the gradient reversal portion back to initial conditions (wash and re-equilibration) of the gradient as part of the same analysis method. The method should end after the gradient has reached and been held at its maximum level for a period of time (we refer to this as the "hold time"). It should NOT switch back to the initial conditions to re-equilibrate the column. Don't make this novice mistake. Column flushing and re-equilibration should be a separate method and/or step, separate from your analysis method. If you include your column flushing (washing) and re-equilibration steps as part of your analysis method, you are also forcing the baseline's slope to change radically. This may interfere with integration plus include extra peaks and baseline changes that you are going to have to integrate, identify and explain to others (e.g. auditors). Additionally, including these wash steps as part of the same method wastes time as you must wait for these steps to complete before you can start to process the data obtained in the analysis method. Summary: Develop methods like the professionals do and create separate Analysis and Flush/Re-Equilibration Methods (or steps).
  7. If your compounds can be seen with a UV/VIS detector, then make sure you are using a Diode Array Detector (aka, DAD or PDA) set to scan a wide range of wavelengths (e.g. 210 - 410 nm). Method development should not be performed using a single or multi-wavelength detector. This invites errors, limits the utility of the method and does not result in any cost savings. You must have a full scanning detector so you can detect all possible peaks at the same time. If you use a single or dual wavelength detector you may not know if an impurity has been introduced or if you have a co-eluter (because you will have no way to detect it). Generic starting settings for Wavelength Bandwidth should be ~ 8 nm, software based Reference Wavelength OFF and the sampling rate must be set to collect at least 20+ peaks/sample peak of the narrowest sample peak observed and integrated (at ½ height). Run with scanning turned 'on' for all analysis methods and review the spectral data for each run. This provides important qualitative data about the compounds which may be used for purity determination and also to demonstrate how the method is selective for the compound(s) of interest. *If your compounds do not absorb well (weak chromophores), then you may still want to have a UV/VIS detector inline (first) with a secondary detector second. The UV/VIS detector will still be very useful for troubleshooting and detecting other compounds. Select an appropriate type of detector and compatible mobile phase as required for the secondary detector (e.g. RID, EC, FLD, ELSD, CAD, MS...). Note: If available, LC-DAD-MS is one of the most useful instrument setups for LC method development.
  8. This needs to be repeated (from Part I)....  Before starting ANY HPLC analysis, the HPLC pump must be running and operating with no problems, achieving a stable baseline, steady flow rate with as little pulsation as possible (~ 1% ripple or less). Accuracy depends on this. Do not begin any HPLC analysis unless the HPLC pumping system is working perfectly.