"Carry-over" is a term used to describe a type of
sample contamination which causes sample peaks to re-appear in later
runs which do not actually contain the sample (e.g. blank runs). The
contamination can last for several sequential runs, often decreasing in
amount after each injection (which is a key observation when
troubleshooting). When proper instrument training has been provided,
modern HPLC system designs make carryover extremely rare, but when it
does appear, the contamination can be due to: (1) A lack of HPLC
maintenance; (2) Overloading samples which foul the column; (3) Poor
Wash Vial Usage and/or Sample Vial Selection; (4) Inadequate operator
training in how to set-up and use the chromatography system. *Note:
Proper operator training greatly reduces the chances of contamination
and is the most overlooked reason for the problem.
The Types of HPLC Carry-Over Contamination; Why They Occur and How To Reduce Them:
(1) A Lack of HPLC Maintenance:
Most auto-injector valves rely on a rotary seal to move the sample from
the needle loop to the flow path of the system. The components within
these valves wear out and should be inspected at least every 6 months
and replaced when needed. Also, always check the needle seat and needle for signs of wear or leaking. Note:
Look for signs of leaks by the injector. Leaks always indicate a
problem and should be fixed immediately. Don't run samples when you have
leaks. Your method and data obtained will be invalid. Any worn parts
should be replaced and the system performance tested. One of the most
common causes of carry-over is due to a worn sample injector valve
rotary seal. A worn seal can allow sample to be retained in the worn grooves, in-between injections, resulting in sample peaks appearing
in subsequent runs. Additionally, buffer salts can lodge between the seals causing leaks or carryover. Routine HPLC service and, if applicable, flushing of all buffers/salts every day can eliminate these issues.
(2) Column Fouling / Overloading of Sample:
If you inject too high a concentration of sample and overload your
column with material, then it is possible that your column will continue
to bleed sample long after the analysis is over. This also happens when the sample has a high affinity for the support you have chosen too. Failure to regularly
flush and clean your HPLC column on a regular basis can also result in a
similar problem as retained material is released from the column over
time. Avoid this problem by performing a loading study to determine how
much material can be effectively loaded on to the column. Next, create a
wash method which utilizes a stronger solvent than your method (often
utilizing a gradient) which will wash away any strongly retained
material in between runs. This is critical if you are running an
isocratic method as material will be retained on
the column and must be washed off at frequent intervals using a stronger
wash solution. *When using only isocratic methods, people often do not initially observe carry-over problems (because the sample just sticks to the column and accumulates over time). When the solvent strength is changed or the method is revised to a gradient, then the problems start... Avoid the problem by selecting the right column (which retains, then elutes ALL of the sample), not overloading the column (do a loading study) and washing the column down with a stronger solution that fully dissolves (not precipitates out) any remaining material off the column after each run.
(3) Wash Vial Usage and/or Sample Vial Selection:
If you are using a modern high-pressure, "Flow-Through" design
autoinjector (e.g. Agilent 1100, 1200, 1260, 1290), then carryover is
rarely an issue as these modern injectors use a high pressure pump to
aspirate and inject the samples directly into the flow path, reducing
the need for any wash stage. The entire HPLC's injection flow path is
continuously washed with mobile phase during the analysis run. This
dramatically reduces the chances of any sample re-appearing in later
runs. The need for a separate wash vial is nearly eliminated in this way
as the needle, needle seat, loop, injector pump and valve are all flushed clean during
each method. Many older auto-injector designs utilize either a low pressure
injector (glass syringe) or injector pump which is not part of the main flow path. These injectors benefit from a separate wash vial as they are
not continuously cleaned. Effective cleaning requires that a wash vial be employed in these cases. It should be filled with mobile
phase or a solution which will dissolve any remaining material which
might still be in the system.
When sticky sample solutions are used, separate Wash
Vials can be used to reduce contamination with either older or newer
injector designs . Sometimes these sticky samples can adhere to the
outside of the needle while it is being withdrawn from a vial which has a
septa which has been punctured many times. High puncture rates tend to
open up the hole resulting in a lack of "wiping' of the needle surface
upon withdrawal. *For vials that are punctured many times, it is
critical to replace the septa OR use septa materials which seal for a
long enough time frame to reduce this effect. Septa needle wiping
eliminates some of this contamination. Two types of contamination can
occur from this problem. (a) When the needle is dipped into a vial (same
or different one) which also has a large septa opening, it can carry
some of the sample with it and deposit it into the new vial (or onto the
septa of the vial). (b) The contamination can also run down the needle
itself and drip onto the needle seat at the time of injection resulting
in contamination of the seat or sample.
One of the easiest
solutions to reduce external needle contamination involves incorporating
a wash vial which contains a solution which is optimized to
quickly dissolve the sample into solution. This sounds simple, but many
chromatographer's choose wash solutions which do not enhance the
cleaning aspect of the needle at all. For example: Mobile phase, which
is normally ideal, but does not work in some cases. Samples such as
peptides, proteins, fats, oils and/or lipids can be troublesome as their
solubility can be at odds with the mobile phase chosen. For the wash
vial to be effective, it must quickly dissolve the material. The needle
can be first "dunked" (dipped) into one vial containing the solution and
withdrawn, followed by an aspiration and wash in a second wash vial. If
needed, you take this cleaning one step further and use additional
aspiration steps to serially dilute any remaining material. These wash
vials must be changed frequently (easily done by having several wash
vial positions programmed in the system). Additionally, the caps should
be left OFF the wash vials to reduce pickup contamination each time they
are used (this step is critical).
Lastly,
if you are analyzing sticky materials which are known to interact with
metals found in chromatography systems, consider using a system which
incorporates bio-compatible materials such as titanium, tantalum and/or
polymers such as PEAK. You can also utilize plastic sample vials (e.g.
PP) or plastic vial inserts too.
(4) Inadequate Operator Training: Good chromatography requires a complete understanding of the hardware used and the fundamentals of HPLC. You must be able to troubleshoot the complete flow path of the system and understand the concepts of chromatography as used in method development. This is not a technique best learned by trial and error, but rather through mentoring using logical steps. Depending
on your skill set, troubleshooting a "carry-over" problem in an HPLC
system can take minutes to months to diagnose and solve. We learn these
skills through hands-on experience and training. Reading many of the
better books and articles on the subject matter helps too. Get as much
practical hands-on training as you can. Ask your supervisor or
manager(s) to invest in you by purchasing professional training for you
in this field so you can learn on your own systems. You will learn far
faster this way and spend less time troubleshooting problems and more
time running samples, accurately in less overall time. Training also
costs just a fraction of what the instrumentation and your salary are.
If you have acquired the fundamental skills, a skilled teacher can
impart about one years worth of practical knowledge to you in as little
as one week of training.
Summary:
The two most common reasons for sample carry-over contamination in an
HPLC or LC/MS system are due to: lack of operator training and/or lack
of system maintenance (most commonly manifested as a worn injector rotor
seal).
Note: This article specifically
addresses carry-over contamination as it relates to the most commonly
used HPLC, UHPLC and LC-MS autoinjectors (or autosampler modules).
You may wish to read a related article on "Troubleshooting HPLC Injectors (Manual and Automated)" found at this link: http://hplctips.blogspot.com/2013/06/troubleshooting-hplc-injectors-manual.html
The HPLC "Hints and Tips" found here will be of interest to anyone performing HPLC, UHPLC (UPLC), LC/MS, SFC, CE or GC analysis. Many of these free tips, taken from professional classes that I teach, discuss fundamental to advanced level chromatography concepts in a practical way. Occasionally, I will also include laboratory computer related tips here too. Please "follow" this blog to receive updates of new posts. Hint: Type keywords into the SEARCH bar to quickly find other articles.
Translator for HPLC HINTS and TIPS for Chromatographers
Saturday, February 28, 2015
Carry-Over (Carryover) Contamination in HPLC and LC-MS Systems
Labels:
Agilent,
Autosampler,
Carryover,
Column,
Contamination,
HP,
HPLC Carry-Over,
Injector,
LC-MS,
LC/MS,
Overloading,
Rotor Seal,
Sample,
Septa,
UHPLC,
UPLC,
Valve,
Wash,
Wash Vial,
Waters
Saturday, January 24, 2015
Useful Windows Command Line Program Shortcuts
Here are some useful Command Line shortcuts which work with many versions of the Microsoft Windows Operating System.
Command Line:
cmd
Advanced Access to User Accounts:
netplwiz
control
userpasswords2
Fix Boot Records:
bootrec
/fixmbr
bootrec
/fixboot
bootrec /rebuildbcd
System Config:
msconfig
Windows Version:
winver
Add Hardware Wizard:
hdwwiz
Control Panel:
control
Device Manager:
devmgmt
Disk Cleanup:
cleanmgr
Display:
dpiscaling
Print Manager:
printmanagement
Windows Explorer:
explorer
Windows Repair Disc:
recdisc
Windows Firewall:
wf
Saturday, December 20, 2014
HPLC Column PORE SIZE (or Pore Diameter) and Retention Time
Think of your typical porous bare silica support as a big sponge full of holes. All of those holes (pores) are where the sample will migrate through before emerging out the other side. With conventional chromatography supports, most of the interaction takes place inside the particle, not on the surface. The size and number of these openings relate to retention time. Besides particle size (particle diameter), pore size is one of the most important characteristics of silica based chromatography supports.
The pore size or pore diameter is often expressed in Angstroms (i.e. 80 A = 8 nm). The degree of porosity relates to the hydrodynamic volume of your sample and is inversely related to the surface area of the support. The larger the surface area of the support (smaller pore size), the longer the possible retention of the sample. For small drug molecule samples under 1,000 daltons (an estimate only) we often use high surface area supports with small pore sizes between 60 and 150 Angstroms (~ 200 to 500 square meters per gram). These provide high retention characteristics useful in separating apart many small compounds in one analysis run. For larger molecules (i.e. peptides and proteins), we employ supports with larger pore sizes (~300 Angstroms). Particles with small pores have larger surface areas which can provide more interaction with the sample. Note: Pore size is often determined using the BET Nitrogen adsorption/desorption equation. Due to endcapping of the support (e.g. C8 or C18), the actual value obtained is often 20-30% less than the original value.
When comparing bare silica columns or trying to identify similar conventional columns for use in a method, pore size must be considered. Manufacturer's publish the pore size in Angstroms (*sometimes in nm) for their different supports. Choosing columns with similar pore sizes is just one of many parameters needed to provide similar retention characteristics.
The pore size or pore diameter is often expressed in Angstroms (i.e. 80 A = 8 nm). The degree of porosity relates to the hydrodynamic volume of your sample and is inversely related to the surface area of the support. The larger the surface area of the support (smaller pore size), the longer the possible retention of the sample. For small drug molecule samples under 1,000 daltons (an estimate only) we often use high surface area supports with small pore sizes between 60 and 150 Angstroms (~ 200 to 500 square meters per gram). These provide high retention characteristics useful in separating apart many small compounds in one analysis run. For larger molecules (i.e. peptides and proteins), we employ supports with larger pore sizes (~300 Angstroms). Particles with small pores have larger surface areas which can provide more interaction with the sample. Note: Pore size is often determined using the BET Nitrogen adsorption/desorption equation. Due to endcapping of the support (e.g. C8 or C18), the actual value obtained is often 20-30% less than the original value.
When comparing bare silica columns or trying to identify similar conventional columns for use in a method, pore size must be considered. Manufacturer's publish the pore size in Angstroms (*sometimes in nm) for their different supports. Choosing columns with similar pore sizes is just one of many parameters needed to provide similar retention characteristics.
Saturday, November 15, 2014
Syringe Filter Selection for HPLC or LC/MS samples
This article will address the use of disposable female, Luer-compatible,
syringe filters without built-in pre-filters for the filtration of individual samples
into vials for HPLC or LC/MS analysis. - Note: 96 or 384 multi-well filtering
plates provide for a better solution when large quantities of samples need to
be filtered. Note: The presented filter membrane material selection criteria also applies to mobile phase filtration too.
The choice of syringe filter depends on the: filter size (volume)
of your sample, the chemical compatibility of the housing and
membrane and desired pore size. Selection of the wrong filter size can result in too much sample
holdup volume (loss of sample on filter) or overloading of the filter (allowing
unfiltered material to pass through). If a membrane or housing is chosen which
is not chemically compatible with your solution, then contamination of the sample
or rupture of the assembly can result. Choosing a filter with too large a pore size can result in material passing through it which could clog or contaminate the solution (i.e. plug an HPLC system or result in a loss of sterility of a solution). Protein binding affinity is another
characteristic of filter membranes and if you are filtering samples of
biological interest, then you will also want to consider this specification in
your selection criteria too (though it will not be discussed in this article).
Syringe Filter Size:
Filters are available in a variety
of sizes which are generally in a disc shape and described by their diameter.
Common sizes available for chromatography samples include: 3 mm, 4 mm, 13 mm and 25 mm (~25 - 30 mm)
diameter discs. The larger the diameter of the disc, the larger the sample
capacity, cross sectional surface area and potential hold-up volume of the
sample on the filter.
Hold-up volume is important because
some of the sample will be retained inside the membrane and/or filter housing. If
too large a filter is selected, samples with small volumes could be lost entirely
in the hold-up volume on the membrane. Smaller filters have lower hold-up volumes. To extract as much sample as possible, be
sure and use a post-filtration air purge to reduce the total hold-up volume.
If the volume of the sample you
wish to filter is under 1 ml, then a 3 mm filter may provide the lowest hold-up
volume and require the smallest amount of solution. To filter samples between
1 ml and 10 ml, the 13 mm diameter filter provides a good balance between hold-up
volume and large filter surface area. Larger sample volumes from 5 ml to 50 ml
are often filtered through the more common 25 mm diameter filters (~4 times the
filtration area as a 13 mm disc).
Chemical Compatibility:
Membrane Material: This is
where you really must consult the manufacturer’s own documentation for the most
compatible filter membrane for both your sample and the solution that will flow
through the filter. To simplify the selection criteria, we can make some generalizations
about some of the different types available:
Cellulose Acetate (CA): Use
with aqueous solutions and a few hydrocarbons only. Low protein binding so good
for many biological samples. Not compatible with ACN or DMSO. Can be
autoclaved.
Nylon: Great general
purpose material and compatible with many HPLC solvents (including THF,
alcohols, ACN), but not strong acids. Nylon has a high affinity to bind
proteins. Can be autoclaved.
Polysulfone / Polyethersulfone
Variants (PS / PES): Commonly used with tissue culture and ion
chromatography samples. Stable with many strong bases and alcohols, but few
HPLC solvents (as it is hydrophilic). Low backpressure and low protein binding.
Not compatible with ACN. Can be autoclaved.
Polypropylene (PP): General
purpose hydrophilic material with resistance to most acids, bases, DMF, DMSO
and alcohols. Not recommended for use with hydrocarbons, esters or solvents
such as ACN. Can be autoclaved.
Polyvinylidene difluoride
(PVDF): Hydrophilic material with broad compatibility. Often a good choice for use
with alcohols, hydrocarbons, biomolecules, ether and ACN. Low protein binding. Can
be autoclaved.
Polytetrafluoroethylene (PTFE): Reported in most brochures to be chemically resistant to almost all solvents, strong acids and bases.
Hydrophobic membrane should be pre-wetted when used with aqueous solutions. Low
protein binding and very strong. Can be autoclaved.
Most chromatography grade syringe filters are constructed of either HDPE or PP. These materials are compatible with a wide range of HPLC solvents and both offer low levels of extractables. HDPE has been reported to be more chemically compatible with aqueous basic solutions of NH4OH than PP.
Most chromatography grade syringe filters are constructed of either HDPE or PP. These materials are compatible with a wide range of HPLC solvents and both offer low levels of extractables. HDPE has been reported to be more chemically compatible with aqueous basic solutions of NH4OH than PP.
Pore Size:
This will depend on your application and a number of different pore sizes are commonly available from vendors (1 micron, 0.8, 0.45 and 0.22 micron are the most common):
For example, is sterilization of the fluid the goal? If so, a 0.22 micron filter is generally accepted as the best choice.
For most chromatography or LC-MS applications either a 0.45 or 0.22 micron filters are preferred.
For example, is sterilization of the fluid the goal? If so, a 0.22 micron filter is generally accepted as the best choice.
For most chromatography or LC-MS applications either a 0.45 or 0.22 micron filters are preferred.
Summary:
- Please refer to the various manufacturers data sheets to select an appropriate syringe filter with: (1) a low hold-up volume; (2) large enough size for the volume of sample; (3) which is chemically compatible with the solution and material you are going to inject through it and (4) lowest protein binding affinity (if applicable).
- To reduce the hold-up volume, use a post-filtration air purge to empty the filter.
- Minimize contamination from extractables (in the plastic) by pre-rinsing the filter membrane with some of the clean solution. This can reduce the amount of detectable extractables in your sample. PTFE based membranes have some of the lowest extractable levels so consider their use if this is an issue.
- If analyte binding is a concern, select one of the membranes which has the lowest binding affinity such as PVDF or PTFE.
Labels:
Chemical,
Choice,
Filter,
Housing,
HPLC,
LC-MS,
LC/MS,
Membrane,
Prep,
PTFE,
Sample,
Sample preparation,
Selection,
Syringe,
Syringe Filter,
UHPLC,
UPLC
Subscribe to:
Comments (Atom)