Saturday, February 1, 2020

Air Bubbles Exiting the HPLC PUMP, Reasons Why.

Reasons For Air Bubbles Exiting The HPLC Pump:

  • Pump Cavitation: When the pump pressure fluctuates wildly up and down, at very low pressures, this is often due to 'pump cavitation'. It is caused by a loss of priming inside the pump (Air, instead of liquid is in the pump's flow path). The HPLC pump should be primed with fresh, degassed mobile phase (following proper procedures) to restore smooth, stable flow. Often, this can be accomplished using the pump, set to a high flow rate, to draw liquid from the bottles. In cases where the pump is not strog enough, manually priming the low pressure lines using a syringe (~ 20 mL) filled with mobile phase and opening (or disconnecting) a fitting at the pump's outlet may aid in priming the system. Note: Depending on the configuration of your HPLC system, to fully prime an HPLC pump, you may need to run 20 or more mLs of solution through EACH channel. Please keep this in mind every time you use the system and every time you prepare or change a mobile phase solution. This article on baseline/pressure fluctuations may assist you in troubleshooting.
  • Loose Connections: If one or more of the low-pressure fittings (nuts and ferrules)  which secure the Teflon tubing to the pump (or vacuum degasser) are damaged or loose, air may enter the system resulting in bubbles. Most pumps use plastic finger-tight style fittings 1/4-28 (or 5/16-24). The threads are soft and can be deformed. When access to these fittings is difficult, sometimes the fittings are left loose and will allow small amounts of air to be drawn in. A build up of salts and/or buffers on the exposed fittings can also allow air into the system (and the presence of deposits on the fittings indicates poor maintenance and a LEAK !). Inspect the tubing and fittings used for proper type, seating depth, wear/condition, cleanliness and/or damage. Replace parts as needed and re-install using the correct amount of torque.
  • Flow Rate Too High, Too Low or Not Enough Degasser Equilibration Time: Degassing efficiency is directly related to the flow rate. Lower flow rates increase the residence time of the mobile phase in the degassing membrane or tubing, improving the gas removal. Higher flow rates provide less time for gas extraction and result in lower degassing efficiency (which equals bubbles in the outlet line). Check with the manufacturer regarding the optimal flow rate range for your degasser to insure you are working  within an acceptable range. Allow enough time for the degasser to reach its set-point and stabilize before use. If the degasser is not operating properly or is unable to "keep up" with the flow rate, then bubbles may be frequently observed in the outlet lines. 
  • Choice of Mobile Phase Liquid: The miscibility of the liquid is also important. If the new mobile phase is not compatible with the previously used mobile phase, pump cavitation may result. Always flush the pump with an intermediate liquid that will dissolve in both the old and new fluids to flush them out before introducing the new mobile phase solution. (such as pure water or IPA, as applicable). The solubility of air (gas) in the specific solution used also affects the efficiency of the vacuum degasser. Aqueous solutions usually hold less gas than popular organic solvents (though air bubbles can be harder to "push" through in water). The amount of dissolved gas inside the liquid relates directly to the time needed to reduce it to acceptable levels for use in HPLC. Be sure to allow enough time to properly degass the new solution.

  • Dirty or Obstructed Solvent Pickup Filters (Bottle filters): Bottle filters should be cleaned or replaced at regular intervals, following routine maintenance SOPs. When they become fouled or obstructed, a vacuum may form as the liquid is drawn into the system. This may result in air being sucked into the tubing or through a fitting (remember that the low pressure Teflon tubing used to connect the bottles to the degasser and pump is porous and allows gas to diffuse through it). The pickup filters should not obstruct the normal flow of solvent (typically they are 10-20 u in porosity). * a quick troubleshooting tip to rule out an obstructed solvent pickup filter is to temporarily remove the filter from the bottle. Observe the back-pressure on the pump to see if it increases and priming is restored. If so, the filter may be clogged. Always replace the filter with a fresh, clean filter and never operate the HPLC without the solvent filters installed.
  • A Sticking Check Valve: The pump's inlet and outlet check valves must function perfectly, all of the time, to maintain proper flow and pump function. If an inlet check valve is not fully closing properly to seal off the high pressures generated inside the pump, then the pump will not be able to maintain pressure or flow. Inspect the check valve. Remove and clean it, per the manufacturer's guidelines (often this involves placing the check valve assembly in a beaker with solvent such as IPA and sonicating for 20 minutes to remove any residues. If cleaning fails to restore proper valve function, then replace the check valve with a new one.

  • Worn Pump Piston Seals (or Pistons): When the piston seals begin to leak, air is allowed into the system. Pump piston seals require regular replacement (they are normal wear items). Scratched or worn pistons may also result in leaks with air getting into the system. Inspect and Test them both for pressure tightness on a scheduled basis or anytime you suspect a problem. Flush the pump with a suitable liquid, then run a high-pressure test to determine if they pass or fail the manufacturer's leak tightness and high pressure tests. Be sure to perform a physical inspection too.

  • Contaminated or Obstructed Pump Outlet Filter: Most HPLC pumps have a small disposable outlet filter installed at or near the pump outlet line (Note: In the case of most Agilent brand HPLC pumps, a small PTFE filter may be found at the outlet valve or inside of the prime-purge valve). These filters should be replaced at regular intervals (monthly is strongly recommended), especially if any aqueous buffers or solutions are used (a they contribute to contamination). Contaminated pump outlet filters may result in a number of pressure instability problems. Abnormally high back-pressure during operation OR when vented to waste are indications it is obstructed. Regular scheduled replacement is the best way to prevent lost time and reduce system contamination.
 Any of the above causes may contribute to valves not functioning properly or air being drawn into the HPLC system. Troubleshooting should begin with the easiest and obvious areas first. Check the condition of the low pressure tubing used to make the connections to and from the mobile phase bottles and pump. If it is kinked, twisted or damaged, replace it with new tubing. Check the fittings used (nuts and ferrules) for tightness and to insure they have been installed properly and are not leaking. Repair all leaks. Keep the system clean (it is easier to monitor and troubleshoot problems when it is clean). Replace any damaged fittings with new ones. Check the solvent pickup filters monthly to insure they are clean and not obstructed. Make sure the flow rate you are using is within the acceptable range for your degasser. 

Has your degasser module been professionally cleaned and serviced within the last 5 years? Are any degasser errors being generated? Is the vacuum degasser making any unusual sounds? Is liquid being emitted from the vacuum pump exhaust port? If any of the answers to these questions are 'yes', then have the HPLC vacuum degasser professionally diagnosed for problems so that repairs can be made to restore function.

Saturday, September 6, 2014

Common Causes of Baseline Noise in HPLC, UHPLC.



Achieving a flat baseline which does not exhibit spikes, ghost peaks, drift or wander in an unpredictable manner should be a primary goal when performing HPLC analysis or developing methods. Methods which result in flat baselines and have well defined, sharp peaks allow for accurate sample area integration. Integration algorithms perform poorly in quantifying peaks on sloped, drifting or noisy baselines. Excessive baseline noise contributes to many problems, including poor quantitation, high %RSD errors, peak identification errors, retention time variation and many other critical problems. Properly developed HPLC methods are reproducible methods which apply and utilize good chromatography fundamentals. Note: "Noise" is a relative term, often w/o meaning. You should always describe it scientifically, measure and compare the signal to noise ration (S/N) of the baseline vs the peak plus note any cyclical patterns (useful in troubleshooting).


Note: A lack of proper training in the operation of the HPLC system, improper start-up or poor quality maintenance of the chromatograph (Examples: failure to degas and purge the system lines before use; poor mixing; an air bubble stuck in a check valve, a bad detector lamp or a leak will often result in baseline noise) are the main causes of noise. Your HPLC system must be optimized for your specific application. Be sure and allow time for the mobile phase to reach full equilibration with the system before starting any analysis. Do not start an analysis until the baseline is stable.

In this article, we will discuss how temperature fluctuations, inadequate mixing, inadequate degassing and flow cell contamination can result in excessive baseline noise. We will provide suggestions on how to reduce or eliminate these problems. Troubleshooting should be done on-site, not over the web or telephone.

TEMPERATURE FLUCTUATIONS:
To obtain reproducible results, the temperature of the HPLC column must be kept constant during each analysis. Laboratory room temperatures often vary by several degrees during the course of one day and these changes will often change the retention characteristics of the sample(s). The 'On' and 'Off' cycling of power from an air conditioner or heating unit will often cause the baseline to drift in a cyclical manner, up and down, during the day (this can often be seen as a clear sine wave pattern when you zoom-in to study the baseline trace over time). Temperature also changes the refractive index of the mobile phase. Light based detectors (UV/VIS, RI...) will show this change as drift up or down). In some cases, a temperature change of plus or minus one degree C from run-to-run can cause changes in retention times which effect reliability of the method. 

To reduce temperature fluctuations, you must control the temperature of the column and mobile phase (if applicable) during the analysis. This is most commonly done by: (a) using equilibrated mobile phase at the start of the day or analysis, (b) keeping the interconnecting lines as short as possible (esp. any which exit the column and go to detectors/flow cells), (c) insulating any stainless steel lines with plastic tubing to reduce heat loss and (d) using a thermostatted column compartment to maintain the column at a single set temperature throughout the day. Control of the column temperature will remove 'temperature' as a variable from your analysis. Temperature should be a constant run to run, not a variable. Be sure and document the temperature selected as part of your method.

INADEQUATE MOBILE PHASE MIXING:
The associated noise and ripple of incomplete mixing can reduce the limit of detection (LOD) and increase integration error. Both high pressure (with separate pumps) and low pressure pumping (one pump with a multi-channel proportioning valve) systems depend on efficient mixing to reduce noise. For gradient analysis, failure to completely mix the mobile phase solution before it enters the HPLC column often results in excessive baseline noise, spikes and poor reproducibility. "Mixing" is often initially accomplished by combining the flow paths of more than one solvent channel together, using a multi-channel gradient valve or tubing. Mixing also performed directly in a mixer installed in the flow path of an HPLC pump. This mixer is often a static mixer (a simple 'Tee', a tube filled with baffles, a frit or beads, valve orifice or microfluidic device) of low volume design for chromatography use, but allows adequate mixing of the liquids within a prescribed flow rate range. The best mixers incorporate longitudinal and radial mixing in-line. A mixer with too low a volume or of insufficient design can result in poor mixing of the mobile phase (note: incorrect solvent compressibility settings can also cause mixing and noise problems too). To reduce mixing problems, first insure that the mobile phases used are fully soluble with each other. Next, make sure that any mixer used is appropriate for the flow rates and volumes you will be using. If needed, run a gradient valve test to insure that each valve channel is working properly, not leaking or introducing any cross-flow leakage to another channel. Monitor the baseline for pressure stability (% ripple), drift and artifacts  (e.g. spikes) in real time to spot problems and make adjustments to correct them. 

INADEQUATE MOBILE PHASE DEGASSING:
For the best results, continuously degas your mobile phase. Reducing the amount of gas will also improve signal to noise levels of detection, reduce drift and reduce pump cavitation. If you are using an electronic vacuum degassing module, make sure it is maintained and working 100%. A faulty degasser may cause more damage (contamination) to your system and methods. Maintain and Repair them just as you do for your other instrument modules. Gas bubbles may cause check valves to malfunction (get stuck), baseline noise spikes to appear randomly, flow rates and/or pressures to become irregular, detector outputs to show high levels of noise (from air in the flow cell) and also cause the loss of prime or cavitation in pumps. To achieve the best balance of low noise levels and high reliability, both aqueous and organic mobile phases should be fully degassed before and during use. This can be accomplished through stand-alone inline vacuum degassing modules or through gentle continuous helium gas sparging (*Helium makes an excellent choice of gas as it is not soluble in the mobile phase. Never use Nitrogen or Argon gas, they are soluble in the liquid!). In all cases, degassing must be continuous (not just done one time). Continuous degassing reduces cyclical noise and signal variations. For this reason, I do not recommend using ultrasonic baths to degas mobile phase solutions as these are not used in a continuous mode. The mobile phase solution starts to re-absorb gas as soon as you stop sonicating the solution. This results in continuous baseline drift (up and down).
Removal of gasses is critical to the function of a modern HPLC pumping system. The liquids used are compressed to very high levels which forces out solubilized gas from the solutions. This is best accomplished before the liquid is transferred into the pump. These gas bubbles must be minimized to achieve desirable baselines. *Even if you use a high pressure pumping system, an inline degassing system reduces the amount of noise and baseline drift. Properly maintain and service your degasser to insure compliant operation. IOW: Whichever method you use, always degas your mobile phase solutions.

FLOW CELLS:
One other less common cause of baseline spikes and random noise is due to either a dirty flow cell (i.e. the windows) or an air bubble trapped inside the flow cell. If the flow cell is suspected of having one of these problems, then it should be carefully rinsed or flushed out with an appropriate mixture of suitable solutions to expel the air bubble or remove the contamination. If possible, keep a spare, 'known good' flow cell on hand to swap out for troubleshooting purposes. This can help to quickly determine where the problem is. This flow cell must be the exact same size and type (volume and path length) for this purpose. If the cell's windows are contaminated and flushing does not restore them, then many manufacturer's offer kits which allow you to replace the windows and gaskets used. Warning: When attempting to clean or repair any flow cell, be sure and work within the manufacturer's operational specifications for the specific flow cell. Some flow cells are not designed to withstand even very low back pressure and damage can result if you exceed their maximum pressure or chemical rating.

Many other types of problems not mentioned in this short article can also cause baseline noise. For example, a sticking inlet or outlet valve on the pump, worn piston seals, worn out detector lamp(s) or detector electrode (EC) can induce noise. In all cases, the cause must be investigated in a logical, step-wise manner. Demonstrate what is working and rule out items one-by-one.

Reference: http://hplctips.blogspot.com/2014/01/diagnosing-troubleshooting-hplc.html