Saturday, October 6, 2018

HPLC UV - VIS Wavelength Accuracy Check (" Calibration ") Notes

To verify correct detector wavelength accuracy of your HPLC UV / VIS module it is periodically necessary to measure the wavelength accuracy against know standards using an appropriate SOP ("fit for purpose"). This may be required as part of a Performance Verification (PV), Installation Qualification (IQ) or Operational Qualification (OQ). 

Wavelength accuracy may be adversely affected (or change) when an UV/VIS detector is serviced/repaired, moved, suffers a physical shock (bumped), large temperature changes occur, a lamp or other optical component is changed, a flow cell is changed, the optics become dirty or contaminated, or due to normal wear and age. The wavelength accuracy of any applicable detectors (e.g. UV, VIS, UV/VIS, DAD, PDA) should be measured on a regular basis as part of "Good Laboratory Practices" (GLP). Depending on the regulations or guidelines applied, most authorities require accuracy to be within 2 to 3 nm of a certified standard within the range used. In practice, we generally achieve accuracy of equal to or better than 0.5 nm across a range of UV / VIS wavelengths. Following good laboratory practice (GLP) requires that we establish the frequency and conditions which determine when they should be verified. Complete documentation of these wavelength checks which describe their purpose, specificity, application and detailed procedures (SOP) should be reviewed.


We present a few suggestions in how to measure the detector wavelength accuracy of your HPLC UV / VIS module. 


  • Built-In Test Methods: Most instrument manufacturers incorporate one or more wavelength accuracy checks directly built into their detectors. This allows quick and accurate measurement of the detector's wavelength accuracy for one or more wavelengths in an automated fashion. Most instruments utilize built-in filters (e.g. holmium oxide) which have been treated with chemicals to provide repeatable wavelength spectra which can be used to determine the accuracy of the detector (and adjust it to within specification in most cases, too). If your instrument has one or more of these built-in test filters, then follow the manufacturer's instructions for using them to measure the wavelength accuracy of your detector. 
  • Using a solution of high purity ANTHRACENE: Dissolved in an HPLC grade alcohol (i.e. Methanol ) or Acetonitrile (for low UV checks), anthracene has a lambda max of 251 nm. A solution concentration of ~ 1 ug / mL for HPLC use can be injected using a standardized method (SOP) and the area% evaluated, one-at-a-time, at several different wavelengths (for VWD or single wavelength detectors) as follows: 249, 250, 251, 252, 253 nm. Relative to the baseline, the areas should show a peak at 251 nm. If you have a scanning UV/VIS detector (aka: DAD or PDA), then you can scan all wavelengths around the 251 nm region and plot the results using just one run to obtain the same type of data.

  • Using a solution of high purity CAFFEINE in HPLC grade water: Caffeine has two useful lambda maximums that we can use for wavelength accuracy checks in the ultraviolet region, 205 nm and 273 nm. We often prepare a range of solutions from 5 ug / mL to 500 ug / mL for linearity testing of UV/VIS detectors, but any of those same solutions could be used for wavelength accuracy checking (similar method as described above for anthracene).

  • One of the most widely used methods requires a solution of HOLMIUM PERCHLORATE  solution (NIST). Available for purchase from many chemical suppliers, this acidic solution provides excellent signals for calibration at well documented transmittance bands (i.e. 241.1, 287.1, 361.5 nm and many others out to ~ 640 nm, depending on the solution it is dissolved in). The detector's flow cell can be filled with the solution and measurements made. The solution is also available coated onto quartz slides and is in fact what is found and used in many detectors today as part of their built-in verification. However, you can still prepare your own test solution.

Notes: A reminder that the solution used to prepare the wavelength check standard(s) in will directly affect the results obtained. If you prepare it in a solution which has strong absorbance at or near the region you test, the results obtained may be inaccurate (e.g. a test std dissolved in MeOH used to measure wavelength accuracy at 205 nm would not be an appropriate choice. A standard dissolved in ethyl acetate would obscure the UV wavelengths below its cutoff of ~ 256 nm). Make sure your SOPs state exactly which solutions are used, how they are prepared and which flow cell are used to make the measurements! Flow cells with different dimensions (i.e. path lengths, volumes) will result in different signal outputs. Different background solutions will also result in different results which can not be directly compared (invalid test). For each test, you must use scientifically appropriate methods and the same conditions to make all measurements.

Saturday, April 21, 2018

The HPLC Restriction Capillary; Troubleshooting, Qualification and Running Without A Column:

Most types of HPLC pumps will not operate properly without 30 or more bars of back-pressure on their outlets to prevent cavitation and excessive pulsation. Columns play a vital role in stabilizing the baseline during an analysis. In this application, they not only aid retention, but act as a cushion or buffer.

When we want to closely replicate the operation of an HPLC system under "normal" conditions and do not want to use an HPLC column in-line (because a column adds variability), we install a "restrictor" such as a restriction capillary in its place. A restriction capillary is often a very narrow ID section of long tubing (capillary) which will restrict the flow of mobile phase through it. For most HPLC systems, a restrictor which is sized to provide about 1,000 to 2,000 psi (~ 70 to 140 Bars) of back-pressure will closely replicate normal operating conditions. The restrictor can be chosen based on length, ID, volume and your flow rate to create this level of back-pressure. You could place a high pressure rated, zero-dead-volume union its place, but in doing so, the system back-pressure may be extremely low ( a few bars) and show poor pump performance. We need to replicate actual analysis conditions during testing or the results obtained may be invalid and unscientific. An HPLC column, with its densely packed small particles inside acts as a pressure pulse buffer and adds a great deal of back-pressure to the HPLC system. That back-pressure greatly improves the stability of the pump operation and overall baseline. HPLC Columns prevents pulsations by acting as a dampener and/or system buffer.

There will be times when you need to operate the HPLC system without an HPLC column installed.

For Example: 
  • Troubleshooting sources of contamination, carryover or artifact peaks on a column;
  • Measuring the HPLC system delay volume (gradient delay);
  • Testing the performance of the injector;
  • Testing the performance of the pump (measure % ripple); 
  • Testing the performance of a detector module (measure S/N);
  • Running HPLC Operational Qualification Tests (OQ);
  • Running HPLC Installation Qualification Tests (IQ);
  • Running Performance Verification Tests on a Module (PV);
  • Running many of the ASTM Tests (e.g. "Baseline Noise & Drift Test").
Example of a commercially available Restriction Capillary (Agilent P/N G1312-67500). You will want to include any needed details of the restriction capillary chosen for your work in the SOP's that you write which utilize it as part of any test (P/N, source, dimensions, volume...).