"Carry-over" is a term used to describe a type of
sample contamination which causes sample peaks to re-appear in later
runs which do not actually contain the sample (e.g. blank runs). The
contamination can last for several sequential runs, often decreasing in
amount after each injection (which is a key observation when
troubleshooting). When proper instrument training has been provided,
modern HPLC system designs make carryover extremely rare, but when it
does appear, the contamination can be due to: (1) A lack of HPLC
maintenance; (2) Overloading samples which foul the column; (3) Poor
Wash Vial Usage and/or Sample Vial Selection; (4) Inadequate operator
training in how to set-up and use the chromatography system. *Note:
Proper operator training greatly reduces the chances of contamination
and is the most overlooked reason for the problem.
The Types of HPLC Carry-Over Contamination; Why They Occur and How To Reduce Them:
(1) A Lack of HPLC Maintenance:
Most auto-injector valves rely on a rotary seal to move the sample from
the needle loop to the flow path of the system. The components within
these valves wear out and should be inspected at least every 6 months
and replaced when needed. Also, always check the needle seat and needle for signs of wear or leaking. Note:
Look for signs of leaks by the injector. Leaks always indicate a
problem and should be fixed immediately. Don't run samples when you have
leaks. Your method and data obtained will be invalid. Any worn parts
should be replaced and the system performance tested. One of the most
common causes of carry-over is due to a worn sample injector valve
rotary seal. A worn seal can allow sample to be retained in the worn grooves, in-between injections, resulting in sample peaks appearing
in subsequent runs. Additionally, buffer salts can lodge between the seals causing leaks or carryover. Routine HPLC service and, if applicable, flushing of all buffers/salts every day can eliminate these issues.
(2) Column Fouling / Overloading of Sample:
If you inject too high a concentration of sample and overload your
column with material, then it is possible that your column will continue
to bleed sample long after the analysis is over. This also happens when the sample has a high affinity for the support you have chosen too. Failure to regularly
flush and clean your HPLC column on a regular basis can also result in a
similar problem as retained material is released from the column over
time. Avoid this problem by performing a loading study to determine how
much material can be effectively loaded on to the column. Next, create a
wash method which utilizes a stronger solvent than your method (often
utilizing a gradient) which will wash away any strongly retained
material in between runs. This is critical if you are running an
isocratic method as material will be retained on
the column and must be washed off at frequent intervals using a stronger
wash solution. *When using only isocratic methods, people often do not initially observe carry-over problems (because the sample just sticks to the column and accumulates over time). When the solvent strength is changed or the method is revised to a gradient, then the problems start... Avoid the problem by selecting the right column (which retains, then elutes ALL of the sample), not overloading the column (do a loading study) and washing the column down with a stronger solution that fully dissolves (not precipitates out) any remaining material off the column after each run.
(3) Wash Vial Usage and/or Sample Vial Selection:
If you are using a modern high-pressure, "Flow-Through" design
autoinjector (e.g. Agilent 1100, 1200, 1260, 1290), then carryover is
rarely an issue as these modern injectors use a high pressure pump to
aspirate and inject the samples directly into the flow path, reducing
the need for any wash stage. The entire HPLC's injection flow path is
continuously washed with mobile phase during the analysis run. This
dramatically reduces the chances of any sample re-appearing in later
runs. The need for a separate wash vial is nearly eliminated in this way
as the needle, needle seat, loop, injector pump and valve are all flushed clean during
each method. Many older auto-injector designs utilize either a low pressure
injector (glass syringe) or injector pump which is not part of the main flow path. These injectors benefit from a separate wash vial as they are
not continuously cleaned. Effective cleaning requires that a wash vial be employed in these cases. It should be filled with mobile
phase or a solution which will dissolve any remaining material which
might still be in the system.
When sticky sample solutions are used, separate Wash
Vials can be used to reduce contamination with either older or newer
injector designs . Sometimes these sticky samples can adhere to the
outside of the needle while it is being withdrawn from a vial which has a
septa which has been punctured many times. High puncture rates tend to
open up the hole resulting in a lack of "wiping' of the needle surface
upon withdrawal. *For vials that are punctured many times, it is
critical to replace the septa OR use septa materials which seal for a
long enough time frame to reduce this effect. Septa needle wiping
eliminates some of this contamination. Two types of contamination can
occur from this problem. (a) When the needle is dipped into a vial (same
or different one) which also has a large septa opening, it can carry
some of the sample with it and deposit it into the new vial (or onto the
septa of the vial). (b) The contamination can also run down the needle
itself and drip onto the needle seat at the time of injection resulting
in contamination of the seat or sample.
One of the easiest
solutions to reduce external needle contamination involves incorporating
a wash vial which contains a solution which is optimized to
quickly dissolve the sample into solution. This sounds simple, but many
chromatographer's choose wash solutions which do not enhance the
cleaning aspect of the needle at all. For example: Mobile phase, which
is normally ideal, but does not work in some cases. Samples such as
peptides, proteins, fats, oils and/or lipids can be troublesome as their
solubility can be at odds with the mobile phase chosen. For the wash
vial to be effective, it must quickly dissolve the material. The needle
can be first "dunked" (dipped) into one vial containing the solution and
withdrawn, followed by an aspiration and wash in a second wash vial. If
needed, you take this cleaning one step further and use additional
aspiration steps to serially dilute any remaining material. These wash
vials must be changed frequently (easily done by having several wash
vial positions programmed in the system). Additionally, the caps should
be left OFF the wash vials to reduce pickup contamination each time they
are used (this step is critical).
Lastly,
if you are analyzing sticky materials which are known to interact with
metals found in chromatography systems, consider using a system which
incorporates bio-compatible materials such as titanium, tantalum and/or
polymers such as PEAK. You can also utilize plastic sample vials (e.g.
PP) or plastic vial inserts too.
(4) Inadequate Operator Training: Good chromatography requires a complete understanding of the hardware used and the fundamentals of HPLC. You must be able to troubleshoot the complete flow path of the system and understand the concepts of chromatography as used in method development. This is not a technique best learned by trial and error, but rather through mentoring using logical steps. Depending
on your skill set, troubleshooting a "carry-over" problem in an HPLC
system can take minutes to months to diagnose and solve. We learn these
skills through hands-on experience and training. Reading many of the
better books and articles on the subject matter helps too. Get as much
practical hands-on training as you can. Ask your supervisor or
manager(s) to invest in you by purchasing professional training for you
in this field so you can learn on your own systems. You will learn far
faster this way and spend less time troubleshooting problems and more
time running samples, accurately in less overall time. Training also
costs just a fraction of what the instrumentation and your salary are.
If you have acquired the fundamental skills, a skilled teacher can
impart about one years worth of practical knowledge to you in as little
as one week of training.
Summary:
The two most common reasons for sample carry-over contamination in an
HPLC or LC/MS system are due to: lack of operator training and/or lack
of system maintenance (most commonly manifested as a worn injector rotor
seal).
Note: This article specifically
addresses carry-over contamination as it relates to the most commonly
used HPLC, UHPLC and LC-MS autoinjectors (or autosampler modules).
You may wish to read a related article on "Troubleshooting HPLC Injectors (Manual and Automated)" found at this link: http://hplctips.blogspot.com/2013/06/troubleshooting-hplc-injectors-manual.html
The HPLC "Hints and Tips" found here will be of interest to anyone performing HPLC, UHPLC (UPLC), LC/MS, SFC, CE or GC analysis. Many of these free tips, taken from professional classes that I teach, discuss fundamental to advanced level chromatography concepts in a practical way. Occasionally, I will also include laboratory computer related tips here too. Please "follow" this blog to receive updates of new posts. Hint: Type keywords into the SEARCH bar to quickly find other articles.
Translator for HPLC HINTS and TIPS for Chromatographers
Saturday, February 28, 2015
Carry-Over (Carryover) Contamination in HPLC and LC-MS Systems
Labels:
Agilent,
Autosampler,
Carryover,
Column,
Contamination,
HP,
HPLC Carry-Over,
Injector,
LC-MS,
LC/MS,
Overloading,
Rotor Seal,
Sample,
Septa,
UHPLC,
UPLC,
Valve,
Wash,
Wash Vial,
Waters
Subscribe to:
Post Comments (Atom)
what is the importance of carry over test while performing calibration
ReplyDelete? why we should perform that?
If you have sample carryover, then your results will be incorrect invalidating the method used and data obtained. *Carryover contributions are often additive to your results (they contribute to the peak area) and may result from poor quality training and/or poor quality method design.
Delete