HPLC UV - VIS Wavelength Accuracy Check (" Calibration ") Notes

To verify correct detector wavelength accuracy of your HPLC UV / VIS module it is periodically necessary to measure the wavelength accuracy against know standards using an appropriate SOP ("fit for purpose"). This may be required as part of a Performance Verification (PV), Installation Qualification (IQ) or Operational Qualification (OQ). 

Wavelength accuracy may be adversely affected (or change) when an UV/VIS detector is serviced/repaired, moved, suffers a physical shock (bumped), large temperature changes occur, a lamp or other optical component is changed, a flow cell is changed, the optics become dirty or contaminated, or due to normal wear and age. The wavelength accuracy of any applicable detectors (e.g. UV, VIS, UV/VIS, DAD, PDA) should be measured on a regular basis as part of "Good Laboratory Practices" (GLP). Depending on the regulations or guidelines applied, most authorities require accuracy to be within 2 to 3 nm of a certified standard within the range used. In practice, we generally achieve accuracy of equal to or better than 0.5 nm across a range of UV / VIS wavelengths. Following good laboratory practice (GLP) requires that we establish the frequency and conditions which determine when they should be verified. Complete documentation of these wavelength checks which describe their purpose, specificity, application and detailed procedures (SOP) should be reviewed.


We present a few suggestions in how to measure the detector wavelength accuracy of your HPLC UV / VIS module. 


  • Built-In Test Methods: Most instrument manufacturers incorporate one or more wavelength accuracy checks directly built into their detectors. This allows quick and accurate measurement of the detector's wavelength accuracy for one or more wavelengths in an automated fashion. Most instruments utilize built-in filters (e.g. holmium oxide) which have been treated with chemicals to provide repeatable wavelength spectra which can be used to determine the accuracy of the detector (and adjust it to within specification in most cases, too). If your instrument has one or more of these built-in test filters, then follow the manufacturer's instructions for using them to measure the wavelength accuracy of your detector. 
  • Using a solution of high purity ANTHRACENE: Dissolved in an HPLC grade alcohol (i.e. Methanol ) or Acetonitrile (for low UV checks), anthracene has a lambda max of 251 nm. A solution concentration of ~ 1 ug / mL for HPLC use can be injected using a standardized method (SOP) and the area% evaluated, one-at-a-time, at several different wavelengths (for VWD or single wavelength detectors) as follows: 249, 250, 251, 252, 253 nm. Relative to the baseline, the areas should show a peak at 251 nm. If you have a scanning UV/VIS detector (aka: DAD or PDA), then you can scan all wavelengths around the 251 nm region and plot the results using just one run to obtain the same type of data.

  • Using a solution of high purity CAFFEINE in HPLC grade water: Caffeine has two useful lambda maximums that we can use for wavelength accuracy checks in the ultraviolet region, 205 nm and 273 nm. We often prepare a range of solutions from 5 ug / mL to 500 ug / mL for linearity testing of UV/VIS detectors, but any of those same solutions could be used for wavelength accuracy checking (similar method as described above for anthracene).

  • One of the most widely used methods requires a solution of HOLMIUM PERCHLORATE  solution (NIST). Available for purchase from many chemical suppliers, this acidic solution provides excellent signals for calibration at well documented transmittance bands (i.e. 241.1, 287.1, 361.5 nm and many others out to ~ 640 nm, depending on the solution it is dissolved in). The detector's flow cell can be filled with the solution and measurements made. The solution is also available coated onto quartz slides and is in fact what is found and used in many detectors today as part of their built-in verification. However, you can still prepare your own test solution.

Notes: A reminder that the solution used to prepare the wavelength check standard(s) in will directly affect the results obtained. If you prepare it in a solution which has strong absorbance at or near the region you test, the results obtained may be inaccurate (e.g. a test std dissolved in MeOH used to measure wavelength accuracy at 205 nm would not be an appropriate choice. A standard dissolved in ethyl acetate would obscure the UV wavelengths below its cutoff of ~ 256 nm). Make sure your SOPs state exactly which solutions are used, how they are prepared and which flow cell are used to make the measurements! Flow cells with different dimensions (i.e. path lengths, volumes) will result in different signal outputs. Different background solutions will also result in different results which can not be directly compared (invalid test). For each test, you must use scientifically appropriate methods and the same conditions to make all measurements.

Saturday, August 11, 2018

Cooling Solutions & Mixtures

An alternative to using an ice bath, peltier cooler or recirculating chiller to cool a solution is to prepare a solution whose freezing point is far below that of water. Here are a few salt mixtures which may be used in the laboratory as "cooling - bath" solutions. 

  • Sodium chloride and crushed ice. * ~ 1 part salt + 3 parts crushed iced.  Good for cooling down to -21C.
  • Potassium chloride and water. * ~ 1 part salt + 4 parts water. Good for cooling down to -11C.
  • Calcium chloride (CaCl2*6H20) and water. *  ~ 2 parts salt + 1.4 parts crushed ice. Good for cooling down to ~ -55C.


 

Saturday, July 7, 2018

HPLC Tubing and Fittings; An introduction to Nuts, Ferrules and Tubing Choices

INTRODUCTION:

Setting up a high pressure liquid chromatography (HPLC) system to run trouble-free takes  patience and a strong set of troubleshooting skills. The patience aspect has usually worn out with most of us, but the troubleshooting skills often come from years of tinkering and practical experience. As a consultant who works with chromatographers on a daily basis, I have found that most chromatographers share many of the same basic HPLC hardware problems. Some of these problems the result of a failure to logically troubleshoot a problem from scratch or by overlooking seemingly minor changes that have been made to the system over time. One common area that is often overlooked in the area of HPLC is that of connection fittings (nuts and ferrules) and tubing selection. Selection and installation of the correct HPLC fittings and tubing can help you avoid future problems while allowing your system to run at peak performance. Common types of high pressure chromatography fittings and tubing found in the laboratory will be discussed in this article.

Please click on this link to download the entire article in PDF format.

Saturday, June 2, 2018

Number of theoretical plates (N), Calculation Formulas


Number of theoretical plates (N):

Often used to quantify the efficiency (performance) of a column (HPLC or GC). "Plates" are expressed per meter of column length and should be calculated based on a retained peak with ideal peak shape or symmetry. 


   N = Plates; tr = Retention Time of Peak; w = Peak width;  w0.5 = Peak width measured at half height.



Two popular formulas are:


Tangent: USP (United States Pharmacopeia / ASTM)

Best for Gaussian peaks. Peak width is often determined at 13.4% of the peak height (w). Inaccurate for peaks which are non-Gaussian, poorly resolved or tail.

   N = 16 (tr / w)2



Half Peak Height: (European Pharmacopeia)

For peaks which are less Gaussian in appearance, using a slightly different formula with the peak width measurement made at the half-height (W0.5). Less accurate for peaks which are poorly resolved or tail.

   N = 5.54 (tr / w0.5)2
 

  • Other formulas, not included, for calculating Plate numbers include: Half Width, Variance Method, Area / Height & Exponential Modified Gaussian (EMG).
  • Caution. HPLC column "Plate" values should not be used for a final determination of efficiency unless you are comparing all results on the same exact HPLC system, which is setup and run under identical conditions each time. Since the result is based on many possible variables, including how your HPLC system is plumbed (dwell volume & tubing ID), the peak's Kprime and symmetry, the detector used, sampling rate, integration quality, flow cell volume, flow rate, actual column used (to name a few), it can easily be manipulated to be very large or small.