HPLC SOLVENT COMPRESSIBILITY - REVISITED

 Twelve years ago I published a short article here (HPLC PUMP SOLVENT COMPRESSIBILITY VALUES) which described the importance of setting the correct solvent compressibility values in the HPLC pump's table. Developing HPLC methods which exhibit smooth, stable baselines, with little measurable signal artifacts (e.g. spikes, noise, oscillation) and minimal pressure fluctuations help insure reliable, repeatable methods. Taking steps to insure that the LC pump operates is setup properly for the method are part of following good chromatography fundamentals

Over the past month I consulted for three different clients who needed help in troubleshooting various "pump stability problems". In all three cases, each HPLC system showed extreme pump pressure cycling, cavitation, noise and instability over time. Pressure fluctuations of 10% (or in one case, 10-30% Ripple values) were observed in several different HPLC methods that were used. One of the very first areas to check for problems with pump pressure instability is mobile phase degassing.  

Proper operation of the HPLC pump requires that efficient degassing of all mobile phases is performed before the liquids enter the pump head. 

Failure to properly degas liquids often results in pump cavitation, check valve sticking and baseline instability. An Inline vacuum degasser or continuous Helium sparing should be used to degas all mobile phase solution for use in HPLC (not sonication or vacuum filtration which perform poorly to solve degassing issues). 

In one of the three cases, the HPLC degasser was found to be broken and long overdue for service. Cleaning and servicing the degasser cured the problem and the method that once showed pressure ripple of >10% now shows no baseline disturbances and very low ripple of ~0.1% at ~ 70 bars system pressure. 

Before I was called in to assist each client, the clients had replaced numerous parts, including: pump seals, check valves, mixers, solvent frits and still had the same baseline instability issues (no change). As recommended by me, two of the clients had their very old degassers cleaned and serviced (as they were long overdue for service), but still had some baseline and pump instability (servicing the degassers improved the baselines, but the pump was not running as it should). In both cases, the cause for the remaining pump instability was quickly identified by me on-site (many problems can be quickly diagnosed on-site).

  • The client had incompatible solvent compressibility values stored as part of their HPLC methods. This resulted in huge baseline disturbances, spikes, cavitation and occasional loss of prime. 
One of the clients normally ran methods containing high percentages of ACN (with some water) for their sample methods, but a few months earlier had switched to running with gradients containing high percentages of methanol. The solvent compressibility values stored in their system were appropriate for WATER, but they never updated them when they used the same method file to run samples in mostly methanol solutions (which need different compressibility values). Though they all had been using HPLC for many years, they had not received basic HPLC instrument training to know how to adjust and optimize these and other important instrument settings for EACH method (they were overwriting each new method, a common new user mistake, when making changes). Once we changed the method's solvent compressibility value to a more compatible one (in their case, for methanol), the baseline smoothed out in just a few minutes and all of the pressure instability issues went away (*they had replaced several thousand dollars worth of perfectly functioning parts trying to solve this issue before I arrived). Professional training in how to use and operate any HPLC instrument should always include how to set and optimize the compressibility value(s). Make sure you know how to incorporate the correct value in each new method that you create. Always spend up-front time to optimize each method for the application before you use it to analyze real samples. The initial time spent getting everything to run smoothly and reliably will improve overall accuracy plus save money and time.
  • Note: In a low-pressure HPLC single-head pumping system with multi-position solvent selector valve (e.g. Most ternary or quaternary systems) one value is allowed, but in a true, dual-head binary pumping system each of the two pump-heads may have a separate field to input the solvent compressibility values.

The importance of inputting the correct and applicable solvent compressibility value(s) into the pump's settings, for each solvent used is one of many steps in creating an optimized HPLC method. There are no universal values, but the instrument manufacturer will have included a generic value in the pump's compressibility settings field. Should you use this generic value?  What are the chances that a randomly selected value used as a 'place holder' in the software is the correct value for your method?  Just as with flow rate, solvent composition, run time, stroke volume, wavelength etc., entering (and saving) the correct solvent compressibility value into EACH method helps to optimize the pumping performance. You will want to select an appropriate value FOR EACH AND EVERY HPLC METHOD YOU CREATE and use (and be sure to save the method with a unique name). Start by loading your HPLC method into the system, then look at the solvent compressibility value(s) used. Are they correct? Change the value(s) shown to values that are appropriate for your method. It is OK to experiment and try different values (we encourage it!). Monitor the S/N levels of the baseline noise for comparison. The instrument manufacturer should provide a table of suggestion solvent compressibility values for use with their system [For HP/Agilent systems, you can see an example table at the link I provided in the first paragraph of this article or review the operator's manual for more information].