In order to reproduce your original method, we must first adjust the flow rate for the new, narrower bore column. The formula to do this is very simple. We decrease the flow rate by using the square of the ratios of the column diameters times the flow rate.
Linear Velocity Change Formula:
( C1 / C2 )2 x original flow rate (ml/min) = new flow rate (ml/min).
Where: C1 = Diameter (mm) of new (smaller) column;
C2 = Diameter (mm) of original column.
------------------------------------------------------------------------------------------------------------------------------
Example #1: Find the new linear flow rate if we use a 2.1 mm ID column in place of a 4.6mm column with an initial flow rate of 1.000 ml/min.
( 2.1 / 4.6 ) 2 x 1.000 = 0.208 (208 ul/min)
Example #2: Find the new linear flow rate if we use a 2.1 mm ID column in place of a 4.6mm column with an initial flow rate of 2.000 ml/min.
( 2.1 / 4.6 ) 2 x 2.000 = 0.416 (416 ul/min)
Example #3: Find the new linear flow rate if we use a 1.0 mm ID column in place of a 4.6mm column with an initial flow rate of 1.500 ml/min.
( 1.0 / 4.6 ) 2 x 1.500 = 0.071 (71 ul/min)
If we assume that the original flow rate is 1.000 ml/min then we can also use this table to get an idea of how the flow rate changes with decreasing column diameter (same particle size and support).
Column I.D. (mm) Flow Rate (ul/min)
4.6 1,000
2.1 208
1.0 47
0.3 4
0.15 1
Summary: Scaling down a method which was originally developed on a 4.6 mm ID column for use on a 2.1 mm ID column (with the same particle size) requires that the flow path of the HPLC system be optimized (reduced) to minimize diffusion and the flow rate reduced five time to achieve the same linear velocity. If the particle size is also going to be reduced from 5u to 2.5u or smaller, then increases in the flow rate may be considered to take advantage of the optimized plate counts using optimized linear velocities (which are much higher for smaller particles).
if we change to higher internal diameter .
ReplyDeleteEx: We are using 4.6mm instead of 3.9mm same formula is applicable or not please let me know
Yes, of course. The formula works for any sizes. That is why I included many examples. Just give it a try.
DeleteThanks it was very useful
ReplyDelete