Proper Wavelength Selection for HPLC Method Development (or Purity Determination)

Selecting the best HPLC wavelength(s) to monitor during an analysis method for use in quantitation and/or purity determination requires both knowledge and careful attention. Here is the basic procedure to use:

Step 1. Create the Method. To determine which UV/VIS detector wavelength(s) should be chosen for the analysis of your sample, you will first need to create a general HPLC Method which retains and resolves the compound(s) of interest on the column (goal is a K prime of >2.0, less than 10.0). Be sure and utilize a scanning diode array detector in full scan mode (often referred to as a photo-diode array detector, PDA or DAD) to scan all relevant wavelengths of your samples (e.g. 210 to 450nm). Note: Your choice of mobile phase and detector settings will effect the S/N values.

Step 2. Determine the lambda max of the sample's spectra using the Data analysis software. Once you have completed the analysis, review the spectral data to determine which prominent peak wavelengths have the maximum signal to noise (S/N) ratio. These “peaks” can be used as the individual wavelengths for integration and purity determination. By sure and double check that any detector options which use a “reference wavelength" are turned ‘OFF’ when running these methods (more info on “reference wavelengths” can be found on this blog in another post). With the wavelength selected, chose an appropriate bandwidth for use (narrow).

Step 3. Edit the HPLC method to use the discreet wavelengths found in step 2. Whenever you run a real sample, continue to use the full scanning mode of the detector so you will know about any other components which absorb at wavelengths far away from and/or near the peak wavelengths. These compounds can add or subtract signal from the main peak making it appear to be more or less concentrated (or more or less pure) than it actually is. If you only monitored the sample with a single wavelength detector, then you would miss this vital information and make errors in your purity or concentration determinations.

Conclusion. (1) Using a multi-wavelength, scanning HPLC detector such as a DAD is one of the most important tools you can use to create accurate and reliable chromatography methods. Always use a scanning DAD for method development to prevent errors. (2) Learning how to correctly use and set up the detector's settings, parameters, special features and options may prevent false or misleading results. Only after you have developed a reliable and repeatable method with good sample retention and peak shape can you begin to report accurate integration and concentration values (and/or make UV/VIS "purity" determinations).

8 comments:

  1. What is we have two samples which use the same wavelength. How do we detect one from the other?

    ReplyDelete
  2. Use good chromatography fundamentals and resolve the two peaks apart first (e.g. baseline resolution > 1.5). This is the goal of good method development. Additionally, be sure and look at the spectral data obtained for each peak. They may be different, and if so, can aid in identification.

    ReplyDelete
  3. On selecting the proper wavelength can I first make A1% conc. by spectrophotometer and determine the lambda max and take this as my wavelength

    ReplyDelete
    Replies
    1. IF your compound is a 100% chemically pure standard, then yes. However, we are normally running HPLC methods because a sample contains a mixture of components. Placing a sample containing a mixture of compounds into the spectrophotometer will result in a mixed spectra, not a pure spectra of one compound. So that would not provide the lamdba max for the sample. As stated above; "To determine which UV/VIS detector wavelength(s) should be chosen for the analysis of your sample, you will first need to create a general HPLC Method which retains the compound(s) of interest on the column".

      Analyzing a mixture in a spectrophotometer cell before developing an HPLC method (or Scanning Diode-Array-Detector...) can provide some useful information. You can read about that right here in a related article, "Method Development Hint: Use your HPLC Diode Array Detector (DAD or PDA) as a Spectrophotometer"; Link: [ http://hplctips.blogspot.com/2013/09/method-development-hint-use-your-hplc.html ].

      Delete
  4. How can i check my detector is ok

    ReplyDelete
    Replies
    1. Run the detector's Operational Qualification (OQ).

      Delete
  5. What does lambda max refer to?? Is it Specific for the compound (through which i can make sure that this peak is of certain compound ) or it refers to the concentration of the compound and will change with the concentration change???

    ReplyDelete
    Replies
    1. "lambda max" is the term which refers to the wavelength of maximum absorbance for the compound. This can be determined by scanning all wavelengths and viewing the resulting spectral data.

      Delete